 IJMPCERO  Vol.8 No.4 , November 2019
Verification of Dosimetric and Positional Accuracy of Dynamic Tumor Tracking Intensity Modulated Radiation Therapy
Abstract:

Purpose: We performed both, dosimetric and positional accuracy verification of dynamic tumor tracking (DTT) intensity modulated radiation therapy (IMRT), with the Vero4DRT system using a moving phantom (QUASAR respiratory motion platform; QUASAR phantom) and system log files. Methods: The QUASAR phantom was placed on a treatment couch. Measurement of the point dose and dose distribution was performed for conventional IMRT, with the QUASAR phantom static and moving; for DTT IMRT, this was performed with the phantom moving for pyramid shaped, prostate, paranasal sinus, and pancreas targets. The QUASAR phantom was driven by a sinusoidal signal in the superior-inferior direction. Furthermore, predicted positional errors induced by the Vero4DRT system and mechanical positional errors of the gimbal head, were calculated using the system log files. Results and Conclusion: For DTT IMRT, the dose at the evaluation point was within 3% compared with the verification plan, and the dose distribution in the passing rates of γ was 97.9%, with the criteria of 3% dose and 3 mm distance to agreement. The position error calculated from the log files was within 2 mm, suggesting the feasibility of employing DTT IMRT with high accuracy using the Vero4DRT system.

Cite this paper: Sueoka, M. , Sawada, A. , Tanabe, H. , Okada, Y. , Taniuchi, S. , Okuuchi, N. , Tanooka, M. , Kokubo, M. and Yamakado, K. (2019) Verification of Dosimetric and Positional Accuracy of Dynamic Tumor Tracking Intensity Modulated Radiation Therapy. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 8, 211-224. doi: 10.4236/ijmpcero.2019.84019.
References

[1]   Matsuo, Y., Shibuya, K., Nakamura, M., Narabayashi, M., Sakanaka, K., Ueki, N., et al. (2012) Dose-Volume Metrics Associated with Radiation Pneumonitis after Stereotactic Body Radiation Therapy for Lung Cancer. International Journal of Radiation Oncology, Biology, Physics, 83, e545-549.
https://doi.org/10.1093/jrr/rrs122

[2]   Matsuo, Y., Onishi, H., Nakagawa, K., Nakamura, M., Ariji, T., Kumazaki, Y., et al. (2013) Guidelines for Respiratory Motion Management in Radiation Therapy. Journal of Radiation Research, 54, 561-568.
https://doi.org/10.1093/jrr/rrs122

[3]   Nakamura, A., Hiraoka, M., Itasaka, S., Nakamura, M., Akimoto, M., Ishihara, Y., et al. (2018) Evaluation of Dynamic Tumor-Tracking Intensity-Modulated Radiotherapy for Locally Advanced Pancreatic Cancer. Scientific Reports, 8, Article No. 17096.
https://doi.org/10.1038/s41598-018-35402-7

[4]   Keall, P.J., Mageras, G.S., Balter, J.M., Emery, R.S., Forster, K.M., Jiang, S.B., et al. (2006) The Management of Respiratory Motion in Radiation Oncology Report of AAPM Task Group 76. Medical Physics, 33, 3874-3900.
https://doi.org/10.1118/1.2349696

[5]   Hirata, Y., Miyamoto, N., Shimizu, M., Yoshida, M., Hiramoto, K., Ichikawa, Y., et al. (2014) International Standardization of Four Dimensional Radiotherapy System: Enhancement of Effects of Irradiation and Assurance of Safety. Synthesiology, 7, 238-246.
https://doi.org/10.5571/synth.7.238

[6]   Kamino, Y., Takayama, K., Kokubo, M., Narita, Y., Hirai, E., Kawawda, N., et al. (2006) Development of a Four-Dimensional Image-Guided Radiotherapy System with a Gimbaled X-Ray Head. International Journal of Radiation Oncology, Biology, Physics, 66, 271-278.
https://doi.org/10.1016/j.ijrobp.2006.04.044

[7]   Mukumoto, N., Nakamura, M., Sawada, A., Takahashi, K., Miyabe, Y., Takayama, K., et al. (2012) Positional Accuracy of Novel X-Ray Image-Based Dynamic Tumor-Tracking Irradiation Using a Gimbaled MV X-Ray Head of a Vero4DRT (MHI-TM2000). Medical Physics, 39, 6287-6296.
https://doi.org/10.1118/1.4754592

[8]   Goto, Y., Ashida, R., Nakamura, A., Itasaka, S., Shibuya, K., Akimoto, M., et al. (2018) Clinical Results of Dynamic Tumor Tracking Intensity-Modulated Radiotherapy with Real-Time Monitoring for Pancreatic Cancers Using a Gimbal Mounted Linac. Oncotarget, 9, 23628-23635.
https://doi.org/10.18632/oncotarget.25310

[9]   Chauffert, B., Mornex, F., Bonnetain, F., Rougier, P., Mariette, C., Bouche, O., et al. (2008) Phase III Trial Comparing Intensive Induction Chemoradiotherapy (60 Gy, Infusional 5-FU and Intermittent Cisplatin) Followed by Maintenance Gemcitabine with Gemcitabine Alone for Locally Advanced Unresectable Pancreatic Cancer. Definitive Results of the 2000-01 FFCD/SFRO Study. Annals of Oncology, 19, 1592-1599.
https://doi.org/10.1093/annonc/mdn281

[10]   Mukumoto, N., Nakamura, M., Yamada, M., Takahashi, K., Akimoto, M., Miyabe, Y., et al. (2016) Development of a Four-Axis Moving Phantom for Patient-Specific QA of Surrogate Signal-Based Tracking IMRT. Medical Physics, 43, 6364.
https://doi.org/10.1118/1.4966130

[11]   Depuydt, T., Verellen, D., Haas, O., Gevaert, T., Linthout, N., Duchateau, M., et al. (2011) Geometric Accuracy of a Novel Gimbals Based Radiation Therapy Tumor Tracking System. Radiotherapy and Oncology, 98, 365-372.
https://doi.org/10.1016/j.radonc.2011.01.015

[12]   Garibaldi, C., Russo, S., Ciardo, D., Comi, S., Seregni, M., Fassi, A., et al. (2015) Geometric and Dosimetric Accuracy and Imaging Dose of the Real-Time Tumour Tracking System of a Gimbal Mounted Linac. Physica Medica, 31, 501-509.
https://doi.org/10.1016/j.ejmp.2015.04.001

[13]   Mukumoto, N., Nakamura, M., Sawada, A., Suzuki, Y., Takahashi, K., Miyabe, Y., et al. (2013) Accuracy Verification of Infrared Marker-Based Dynamic Tumor-Tracking Irradiation Using the Gimbaled X-Ray Head of the Vero4DRT (MHI-TM2000). Medical Physics, 40, Article ID: 041706.
https://doi.org/10.1118/1.4794506

[14]   Tatsumi, D., Hosono, M.N., Nakada, R., Ishii, K., Tsutsumi, S., Inoue, M., et al. (2011) Direct Impact Analysis of Multi-Leaf Collimator Leaf Position Errors on Dose Distributions in Volumetric Modulated Arc Therapy: A Pass Rate Calculation between Measured Planar Doses with and without the Position Errors. Physics in Medicine and Biology, 56, N237-N246.
https://doi.org/10.1088/0031-9155/56/20/N03

[15]   Japan Society for Radiation Oncology Quality Assurance Committee (2004) Guideline about Securing the Mechanical Precision of Intensity-Modulated Radiation Therapy Using Multi-Leaf Collimator (Ver. 1). The Official Journal of the Japanese Society for Therapeutic Radiology and Oncology, 16, 197-203.

[16]   Low, D.A., Harms, W.B., Mutic, S. and Purdy, J.A. (1998) A Technique for the Quantitative Evaluation of Dose Distributions. Medical Physics, 25, 656-661.
https://doi.org/10.1118/1.598248

[17]   Kito, S. (2014) Actual Practices of Dynamic Tumor Tracking Irradiation at Komagome Hospital. Journal of Radiotherapy Section of Japanese Society of Radiological Technology, 28, 43-48. (In Japanese)

[18]   Li, X., Yang, Y., Li, T., Fallon, K., Heron, D.E. and Huq, M.S. (2013) Dosimetric Effect of Respiratory Motion on Volumetric-Modulated Arc Therapy-Based Lung SBRT Treatment Delivered by TrueBeam Machine with Flattening Filter-Free Beam. Journal of Applied Clinical Medical Physics, 14, 4370.
https://doi.org/10.1120/jacmp.v14i6.4370

[19]   Chui, C.S., Yorke, E. and Hong, L. (2003) The Effects of Intra-Fraction Organ Motion on the Delivery of Intensity-Modulated Field with a Multileaf Collimator. Medical Physics, 30, 1736-1746.
https://doi.org/10.1118/1.1578771

[20]   Gierga, D.P., Chen, G.T., Kung, J.H., Betke, M., Lombardi, J. and Willett, C.G. (2004) Quantification of Respiration-Induced Abdominal Tumor Motion and Its Impact on IMRT Dose Distributions. International Journal of Radiation Oncology, Biology, Physics, 58, 1584-1595.
https://doi.org/10.1016/j.ijrobp.2003.09.077

[21]   Bortfeld, T., Jokivarsi, K., Goitein, M., Kung, J. and Jiang, S.B. (2002) Effects of Intra-Fraction Motion on IMRT Dose Delivery: Statistical Analysis and Simulation. Physics in Medicine and Biology, 47, 2203-2220.
https://doi.org/10.1088/0031-9155/47/13/302

[22]   Yang, J.N., Mackie, T.R., Reckwerdt, P., Deasy, J.O. and Thomadsen, B.R. (1997) An Investigation of Tomotherapy Beam Delivery. Medical Physics, 24, 425-436.
https://doi.org/10.1118/1.597909

[23]   Jiang, S.B., Pope, C., Al Jarrah, K.M., Kung, J.H., Bortfeld, T. and Chen, G.T. (2003) An Experimental Investigation on Intra-Fractional Organ Motion Effects in Lung IMRT Treatments. Physics in Medicine and Biology, 48, 1773-1784.
https://doi.org/10.1088/0031-9155/48/12/307

[24]   Seco, J., Sharp, G.C., Turcotte, J., Gierga, D., Bortfeld, T. and Paganetti, H. (2007) Effects of Organ Motion on IMRT Treatments with Segments of Few Monitor Units. Medical Physics, 34, 923-934.
https://doi.org/10.1118/1.2436972

[25]   Ehler, E.D., Nelms, B.E. and Tome, W.A. (2007) On the Dose to a Moving Target While Employing Different IMRT Delivery Mechanisms. Radiotherapy and Oncology, 83, 49-56.
https://doi.org/10.1016/j.radonc.2007.02.007

[26]   Kang, H., Yorke, E.D., Yang, J., Chui, C.S., Rosenzweig, K.E. and Amols, H.I. (2010) Evaluation of Tumor Motion Effects on Dose Distribution for Hypofractionated Intensity-Modulated Radiotherapy of Non-Small-Cell Lung Cancer. Journal of Applied Clinical Medical Physics, 11, 3182.
https://doi.org/10.1120/jacmp.v11i3.3182

[27]   Colvill, E., Booth, J., Nill, S., Fast, M., Bedford, J., Oelfke, U., et al. (2016) A Dosimetric Comparison of Real-Time Adaptive and Non-Adaptive Radiotherapy: A Multi-Institutional Study Encompassing Robotic, Gimbaled, Multileaf Collimator and Couch Tracking. Radiotherapy and Oncology, 119, 159-165.
https://doi.org/10.1016/j.radonc.2016.03.006

[28]   Ezzell, G.A., Burmeister, J.W., Dogan, N., LoSasso, T.J., Mechalakos, J.G., Mihailidis, D., et al. (2009) IMRT Commissioning: Multiple Institution Planning and Dosimetry Comparisons, a Report from AAPM Task Group 119. Medical Physics, 36, 5359-5373.
https://doi.org/10.1118/1.3238104

[29]   Alber, M., Broggi, S., Wagter, C.D., Eichwurzel, I., Engstrom, P., Fiorino, C., et al. (2008) Guidelines for the Verifi Cation of IMRT. ESTRO Booklet No. 9.

 
 
Top