Periodic Measurements of Passive Proton Beam Width Using Radiochromic Film in Fixed Gantry System
Abstract:
Background and Aim: Irradiation methods such as double scattering method and spot scanning method have been used in proton beam treatment devices. In the scattering method, a ridge filter or a range modulation wheel is used to create a spread-out Bragg peak, but the distribution at the patient position may change due to positional deviation of the incident beam. Therefore, assessment of the incident position of the beam is very important even in the scattering method. To investigate the width and distribution of the proton beam before entering the RMW, a radiochromic film was installed at the outlet of the transport pipe and the entrance of the profile-monitoring detector. Methods: In this study, the distributions of the beam at the exit of the transport pipe and the entrance of the monitor detector were measured using films. The beam width was measured from the full width at half maximum of the profile obtained from the distribution. Measurements were conducted every month for 10 months. Results: Beams of widths ranging from 1.82 to 2.30 mm in the horizontal direction and 4.25 to 5.33 mm in the vertical direction were outputted from the exit of the transport pipe. Beams of widths ranging from 2.16 to 2.67 mm in the horizontal direction and 4.06 to 5.31 mm in the vertical direction were outputted from the entrance of the monitor detector. The maximum width fluctuation for 10 months was 0.55 mm in the horizontal direction and 1.26 mm in the vertical direction at the entrance of the monitor detector. Conclusions: The distribution was obtained before the proton beam was scattered by the scatterer, and then we propose a method to periodically measure and monitor the changes in the beam distributions every month.
Cite this paper: Shimozato, T. , Yasui, K. , Kinou, H. and Komatsu, F. (2019) Periodic Measurements of Passive Proton Beam Width Using Radiochromic Film in Fixed Gantry System. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 8, 193-203. doi: 10.4236/ijmpcero.2019.84017.
References

[1]   De Laney, T.F. and Kooy, H.M. (2008) Proton and Charged Particle Radiotherapy. Lippincott Williams & Wilkins, Philadelphia, 33-49.

[2]   ICRU (2007) Prescribing, Recording, and Reporting Proton-Beam Therapy. ICRU Report No. 78. International Commission on Radiation Units and Measurements, Bethesda.

[3]   Kanai, T., Kawachi, K., Kumamoto, Y., et al. (1980) Spot Scanning System for Proton Radiotherapy. Medical Physics, 7, 365-369. https://doi.org/10.1118/1.594693

[4]   Titt, U., Mirkovic, D., Sawakuchi, G.O., et al. (2010) Adjustment of the Lateral and Longitudinal Size of Scanned Proton Beam Spots Using a Pre-Absorber to Optimize Penumbrae and Delivery Efficiency. Physics in Medicine and Biology, 55, 7097-7106.
https://doi.org/10.1088/0031-9155/55/23/S10

[5]   Polf, J.C., Harvey, M.C., Titt, U., Newhauser, W.D. and Smith, A.R. (2007) Initial Beam Size Study for Passive Scatter Proton Therapy. I. Monte Carlo Verification. Medical Physics, 34, 4213-4218. https://doi.org/10.1118/1.2767934

[6]   Polf, J.C., Harvey, M.C. and Smith, A.R. (2007) Initial Beam Size Study for Passive Scatter Proton Therapy. II. Changes in Delivered Depth Dose Profiles. Medical Physics, 34, 4219-4222. https://doi.org/10.1118/1.2789497

[7]   Toshito, T., Omachi, C., Kibe, Y., et al. (2016) A Proton Therapy System in Nagoya Proton Therapy Center. Australasian Physical and Engineering Sciences in Medicine, 39, 645-654. https://doi.org/10.1007/s13246-016-0456-8

[8]   Ferreira, B.C., Lopes, M.C. and Capela, M. (2009) Evaluation of an Epson Flatbed Scanner to Read Gafchromic EBT Films for Radiation Dosimetry. Physics in Medicine and Biology, 54, 1073-1085. https://doi.org/10.1088/0031-9155/54/4/017

 
 
Top