 IJMPCERO  Vol.8 No.2 , May 2019
Treatment-Plan Comparison of Three Advanced Radiation Treatment Modalities for Fractionated Stereotactic Radiotherapy to the Head and Neck
Abstract:
Purpose: Fractionated stereotactic radiosurgery (FSRS) can be given with at least three modalities: Gamma Knife, with the noninvasive frameless extend system (GKE); linear accelerator-based volumetric modulated arc therapy (VMAT); and intensity-modulated proton therapy (IMPT). We extracted treatment plans for patients who had received FSRS with GKE for recurrent skull base or intracranial tumors, created corresponding plans for VMAT and IMPT, and compared the quality of the three sets of plans. Methods and materials: Plans were extracted for 9 patients with recurrent malignant skull-base tumors (n = 6) or large intracranial tumors (n = 3) who had received FSRS with GKE (median dose 24 Gy in 3 fractions) in 2013 through 2015. Plans for VMAT were generated with a TrueBeam STx LINAC machine using 6-MV photons, and plans for IMPT were generated with multi-field optimization. The optimized VMAT and IMPT plans were normalized to achieve the best possible target coverage while meeting the same dose-volume constraints on organs at risk (OARs) as the GKE plans. Plans were evaluated on the basis of target coverage, conformity index, homogeneity index, gradient index, and treatment efficiency. Results: The median target volume was 10.2 cm3 (range 1.9 - 33.8 cm3). The VMAT and IMPT plans met all OAR constraints, and target coverage and conformity were comparable among all plans. VMAT and IMPT plans showed significantly better target uniformity and treatment delivery efficiency (P < 0.001). The gradient index and low-dose-bath were superior in the GKE plans (P < 0.001), indicating smaller irradiated volumes. When inverse planning was used, VMAT plans could achieve a similar or steeper dose drop-off at the target-OAR boundary than GKE plans. Conclusion: FSRS for skull base and large intracranial lesions delivered by VMAT and IMPT can achieve comparable target coverage, conformity, and sparing of critical structure as the GKE while providing superior target uniformity and treatment delivery efficiency. The GKE had superior high-dose gradients outside the target and thus better protected surrounding normal structures.
Cite this paper: Wang, H. , Yang, J. , Zhang, X. , Li, J. , Frank, S. , Zhao, Z. , Luo, D. , Zhu, X. , Wang, C. , Tung, S. , Garden, A. , Rosenthal, D. , Fuller, C. , Gunn, G. , Ghia, A. , Reddy, J. , Raza, S. , Monte, F. , Chambers, M. , Brown, P. , Su, S. and Phan, J. (2019) Treatment-Plan Comparison of Three Advanced Radiation Treatment Modalities for Fractionated Stereotactic Radiotherapy to the Head and Neck. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 8, 106-120. doi: 10.4236/ijmpcero.2019.82010.
References

[1]   Mack, A., Czempiel, H., Kreiner, H.-J., et al. (2002) Quality Assurance in Stereotactic Space. A System Test for Verifying the Accuracy of Aim in Radiosurgery. Medical Physics, 29, 561-568.
https://doi.org/10.1118/1.1463062

[2]   Novotny, J., Bhatnagar, J.P., Xu, Y., et al. (2014) Long-Term Stability of the Leksell Gamma Knife® PerfexionTM Patient Positioning System (PPS). Medical Physics, 41, Article ID: 031711.
https://doi.org/10.1118/1.4866225

[3]   Ma, L., Chiu, J., Hoye, J., et al. (2014) Quality Assurance of Stereotactic Alignment and Patient Positioning Mechanical Accuracy for Robotized Gamma Knife Radiosurgery. Physics in Medicine and Biology, 59, N221. https://doi.org/10.1088/0031-9155/59/23/N221

[4]   Ma, L., Chuang, C., Descovich, M., et al. (2008) Whole-Procedure Clinical Accuracy of Gamma Knife Treatments of Large Lesions. Medical Physics, 35, 5110-5114.
https://doi.org/10.1118/1.2987669

[5]   Karaiskos, P., Moutsatsos, A., Pappas, E., et al. (2014) A Simple and Efficient Methodology to Improve Geometric Accuracy in Gamma Knife Radiation Surgery: Implementation in Multiple Brain Metastases. International Journal of Radiation Oncology, Biology, Physics, 90, 1234-1241.
https://doi.org/10.1016/j.ijrobp.2014.08.349

[6]   Kress, M.-A.S., Sen, N., Unger, K.R., et al. (2015) Safety and Efficacy of Hypofractionated Stereotactic Body Reirradiation in Head and Neck Cancer: Long-Term Follow-Up of a Large Series. Head & Neck, 37, 1403-1409. https://doi.org/10.1002/hed.23763

[7]   Tanyi, J.A., Summers, P.A., McCracken, C.L., et al. (2009) Implications of a High-Definition Multileaf Collimator (HD-MLC) on Treatment Planning Techniques for Stereotactic Body Radiation Therapy (SBRT): A Planning Study. Radiation Oncology, 4, 22.
https://doi.org/10.1186/1748-717X-4-22

[8]   Smith, A.R. (2009) Vision 20/20: Proton Therapy. Medical Physics, 36, 556-568.
https://doi.org/10.1118/1.3058485

[9]   Chhabra, A., Langen, K. and Mehta, M.P. (2016) An Overview of Modern Proton Therapy. Chinese Clinical Oncology, 5, 48. https://doi.org/10.21037/cco.2016.05.06

[10]   Apinorasethkul, O., Kirk, M., Teo, K., et al. (2017) Pencil Beam Scanning Proton Therapy vs. Rotational Arc Radiation Therapy: A Treatment Planning Comparison for Postoperative Oropharyngeal Cancer. Medical Dosimetry, 42, 7-11.
https://doi.org/10.1016/j.meddos.2016.09.004

[11]   Steneker, M., Lomax, A. and Schneider, U. (2006) Intensity Modulated Photon and Proton Therapy for the Treatment of Head and Neck Tumors. Radiotherapy and Oncology, 80, 263-267. https://doi.org/10.1016/j.radonc.2006.07.025

[12]   van de Water, T.A., Lomax, A.J., Bijl, H.P., et al. (2012) Using a Reduced Spot Size for Intensity-Modulated Proton Therapy Potentially Improves Salivary Gland-Sparing in Oropharyngeal Cancer. International Journal of Radiation Oncology, Biology, Physics, 82, e313-e319. https://doi.org/10.1016/j.ijrobp.2011.05.005

[13]   Welsh, J., Gomez, D., Palmer, M.B., et al. (2011) Intensity-Modulated Proton Therapy Further Reduces Normal Tissue Exposure during Definitive Therapy for Locally Advanced Distal Esophageal Tumors: A Dosimetric Study. International Journal of Radiation Oncology, Biology, Physics, 81, 1336-1342. https://doi.org/10.1016/j.ijrobp.2010.07.2001

[14]   Liu, M., Snyder, K., Zhao, B., et al. (2015) In VMAT of Spine Stereotactic Radiosurgery a 1 Mm Grid Size Increases Dose Gradient and Lowers Cord Dose Significantly Relative to a 2.5 Mm Grid Size. Medical Physics, 42, 3446. https://doi.org/10.1118/1.4924849

[15]   Gillin, M.T., Sahoo, N., Bues, M., et al. (2010) Commissioning of the Discrete Spot Scanning Proton Beam Delivery System at the University of Texas M.D. Anderson Cancer Center, Proton Therapy Center, Houston. Medical Physics, 37, 154-163. https://doi.org/10.1118/1.3259742

[16]   Holliday, E.B. and Frank, S.J. (2016) Proton Therapy for Nasopharyngeal Carcinoma. Chinese Clinical Oncology, 5, 25. https://doi.org/10.21037/cco.2016.03.05

[17]   Kohno, R., Cao, W., Yepes, P., et al. (2017) Dose Comparison between Eclipse Dose Calculation and Fast Dose Calculator in Single- and Multi-Field Optimization Intensity-Modulated Proton Therapy Plans with Various Multi-Beams for Brain Cancer. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 6, 12.

[18]   Paddick, I. (2000) A Simple Scoring Ratio to Index the Conformity of Radiosurgical Treatment Plans. Technical Note. Journal of Neurosurgery, 93, 219-222.
https://doi.org/10.3171/jns.2000.93.supplement_3.0219

[19]   Suzuki, K., Palmer, M.B., Sahoo, N., et al. (2016) Quantitative Analysis of Treatment Process Time and Throughput Capacity for Spot Scanning Proton Therapy. Medical Physics, 43, 3975-3986. https://doi.org/10.1118/1.4952731

[20]   Sayer, F.T., Sherman, J.H., Yen, C.P., et al. (2011) Initial Experience with the eXtend System: A Relocatable Frame System for Multiple-Session Gamma Knife Radiosurgery. World Neurosurgery, 75, 665-672. https://doi.org/10.1016/j.wneu.2010.12.051

[21]   Schlesinger, D., Xu, Z., Taylor, F., et al. (2012) Interfraction and Intrafraction Performance of the Gamma Knife Extend System for Patient Positioning and Immobilization. Special Supplements, 117, 217-224. https://doi.org/10.3171/2012.6.GKS12989

[22]   Ruschin, M., Nayebi, N., Carlsson, P., et al. (2010) Performance of a Novel Repositioning Head Frame for Gamma Knife Perfexion and Image-Guided Linac-Based Intracranial Stereotactic Radiotherapy. International Journal of Radiation Oncology, Biology, Physics, 78, 306-313.
https://doi.org/10.1016/j.ijrobp.2009.11.001

[23]   Clivio, A., Vanetti, E., Rose, S., et al. (2015) Evaluation of the Machine Performance Check Application for TrueBeam Linac. Radiation Oncology, 10, 97.
https://doi.org/10.1186/s13014-015-0381-0

[24]   Murphy, M.J. (2009) Intrafraction Geometric Uncertainties in Frameless Image-Guided Radiosurgery. International Journal of Radiation Oncology, Biology, Physics, 73, 1364-1368. https://doi.org/10.1016/j.ijrobp.2008.06.1921

[25]   McDonald, D., Schuler, J., Takacs, I., et al. (2014) Comparison of Radiation Dose Spillage from the Gamma Knife Perfexion with That from Volumetric Modulated Arc Radiosurgery during Treatment of Multiple Brain Metastases in a Single Fraction. Journal of Neurosurgery, 121, 51-59. https://doi.org/10.3171/2014.7.GKS141358

[26]   Thomas, E.M., Popple, R.A., Wu, X., et al. (2014) Comparison of Plan Quality and Delivery Time between Volumetric Arc Therapy (RapidArc) and Gamma Knife Radiosurgery for Multiple Cranial Metastases. Neurosurgery, 75, 409-417. https://doi.org/10.1227/NEU.0000000000000448

[27]   Kandula, S., Zhu, X., Garden, A.S., et al. (2013) Spot-Scanning Beam Proton Therapy vs Intensity-Modulated Radiation Therapy for Ipsilateral Head and Neck Malignancies: A Treatment Planning Comparison. Medical Dosimetry, 38, 390-394.
https://doi.org/10.1016/j.meddos.2013.05.001

[28]   van de Water, T.A., Lomax, A.J., Bijl, H.P., et al. (2011) Potential Benefits of Scanned Intensity-Modulated Proton Therapy versus Advanced Photon Therapy with Regard to Sparing of the Salivary Glands in Oropharyngeal Cancer. International Journal of Radiation Oncology, Biology, Physics, 79, 1216-1224. https://doi.org/10.1016/j.ijrobp.2010.05.012

[29]   Bokrantz, R. and Fredriksson, A. (2017) Scenario-Based Radiation Therapy Margins for Patient Setup, Organ Motion, and Particle Range Uncertainty. Physics in Medicine and Biology, 62, 1342-1357. https://doi.org/10.1088/1361-6560/aa524d

[30]   Paganetti, H. (2012) Range Uncertainties in Proton Therapy and the Role of Monte Carlo Simulations. Physics in Medicine and Biology, 57, R99-R117.
https://doi.org/10.1088/0031-9155/57/11/R99

[31]   Warren, S., Partridge, M., Bolsi, A., et al. (2016) An Analysis of Plan Robustness for Esophageal Tumors: Comparing Volumetric Modulated Arc Therapy Plans and Spot Scanning Proton Planning. International Journal of Radiation Oncology, Biology, Physics, 95, 199-207.
https://doi.org/10.1016/j.ijrobp.2016.01.044

[32]   Grant, J.D. and Chang, J.Y. (2014) Proton-Based Stereotactic Ablative Radiotherapy in Early-Stage Non-Small-Cell Lung Cancer. Biomedical Research International, 2014, Article ID: 389048. https://doi.org/10.1155/2014/389048

[33]   Register, S.P., Zhang, X., Mohan ,R., et al. (2011) Proton Stereotactic Body Radiation Therapy for Clinically Challenging Cases of Centrally and Superiorly Located Stage I Non-Small-Cell Lung Cancer. International Journal of Radiation Oncology, Biology, Physics, 80, 1015-1022.
https://doi.org/10.1016/j.ijrobp.2010.03.012

[34]   Schuemann, J., Dowdell, S., Grassberger, C., et al. (2014) Site-Specific Range Uncertainties Caused by Dose Calculation Algorithms for Proton Therapy. Physics in Medicine and Biology, 59, 4007-4031. https://doi.org/10.1088/0031-9155/59/15/4007

[35]   Unkelbach, J., Bortfeld, T., Martin, B.C., et al. (2009) Reducing the Sensitivity of IMPT Treatment Plans to Setup Errors and Range Uncertainties via Probabilistic Treatment Planning. Medical Physics, 36, 149-163. https://doi.org/10.1118/1.3021139

[36]   Wang, H., Wang, C., Tung, S., et al. (2016) Improved Setup and Positioning Accuracy Using a Three-Point Customized Cushion/Mask/Bite-Block Immobilization System for Stereotactic Reirradiation of Head and Neck Cancer. Journal of Applied Clinical Medical Physics, 17, 6038.
https://doi.org/10.1120/jacmp.v17i3.6038

[37]   Kocak-Uzel, E., Gunn, G.B., Colen, R.R., et al. (2014) Beam Path Toxicity in Candidate Organs-at-Risk: Assessment of Radiation Emetogenesis for Patients Receiving Head and Neck Intensity Modulated Radiotherapy. Radiotherapy and Oncology, 111, 281-288.
https://doi.org/10.1016/j.radonc.2014.02.019

 
 
Top