Back
 IJMPCERO  Vol.7 No.4 , November 2018
Application of Variance Component Analysis (ANOVA) in Setup Errors and PTV Margins for Lung Cancer with Stereotactic Body Radiation Therapy (SBRT)
Abstract: Purpose: To investigate the feasibility of applying ANOVA newly proposed by Yukinori to verify the setup errors, PTV (Planning Target Volume) margins, DVH for lung cancer with SBRT. Methods: 20 patients receiving SBRT to 50 Gy in 5 fractions with a Varian iX linear acceleration were selected. Each patient was scanned with kV-CBCT before the daily treatment to verify the setup position. Two other error calculation methods raised by Van Herk and Remeijer were also compared to discover the statistical difference in systematic errors (Σ), random errors (σ), PTV margins and DVH. Results: Utilizing two PTV margin calculation formulas (Stroom, Van Herk), PTV calculated by Yukinori method in three directions were (5.89 and 3.95), (5.54 and 3.55), (3.24 and 0.78) mm; Van Herk method were (6.10 and 4.25), (5.73 and 3.83), (3.51 and 1.13) mm; Remeijer method were (6.39 and 4.57), (5.98 and 4.10), (3.69 and 1.33) mm. The volumes of PTV using Yukinori method were significantly smaller (P < 0.05) than Van Herk method and Remeijer method. However, dosimetric indices of PTV (D98, D50, D2) and for OARs (Mean Dose, V20, V5) had no significant difference (P > 0.05) among three methods. Conclusions: In lung SBRT treatment, due to fraction reduction and high level of dose per fraction, ANOVA was able to offset the effect of random factors in systematic errors, reducing the PTV margins and volumes. However, no distinct dose distribution improvement was founded in target volume and organs at risk.
Cite this paper: Huang, X. , Zhang, J. , Xie, C. , Zhou, Y. and Quan, H. (2018) Application of Variance Component Analysis (ANOVA) in Setup Errors and PTV Margins for Lung Cancer with Stereotactic Body Radiation Therapy (SBRT). International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 7, 522-538. doi: 10.4236/ijmpcero.2018.74044.
References

[1]   Timmerman, R., Kavanagh, B., Cho, L., Papiez, L. and Xing, L. (2007) Stereotactic Body Radiation Therapy in Multiple Organ Sites. Journal of Clinical Oncology, 25, 947-952.
https://doi.org/10.1200/JCO.2006.09.7469

[2]   Alamri, I., Faria, S., Gluszko, J. and Patrocinio, H. (2016) Intra-Fraction Tumor Position Assessment for Lung SBRT in Patients Treated without Customized Immobilization Devices. Medical Physics, 43, 4947.
https://doi.org/10.1118/1.4961820

[3]   Jiang, R., Zhan, L. and Osei, E. (2015) Lung SBRT Dosimetric Comparison of 3D Conformal and RapidArc Planning. Medical Physics, 42, 3459.
https://doi.org/10.1118/1.4924900

[4]   Ong, C., Palma, D., Verbakel, W., Slotman, B. and Senan, S. (2010) Treatment of Large Stage I-II Lung Tumors Using Stereotactic Body Radiotherapy (SBRT): Planning Considerations and Early Toxicity. Radiotherapy and Oncology, 97, 431-436.
https://doi.org/10.1016/j.radonc.2010.10.003

[5]   Van Herk, M. (2004) Errors and Margins in Radiotherapy. Seminars in Radiation Oncology, 14, 52-64.
https://doi.org/10.1053/j.semradonc.2003.10.003

[6]   Remeijer, P., Geerlof, E., Ploeger, L., Gilhuijs, K., van Herk, M. and Lebesque, J.V. (2000) 3-D Portal Image Analysis in Clinical Practice: An Evaluation of 2-D and 3-D Analysis Techniques as Applied to 30 Prostate Cancer Patients. International Journal of Radiation Oncology, Biology, Physics, 46, 1281-1290.
https://doi.org/10.1016/S0360-3016(99)00468-X

[7]   Gordon, J. and Siebers, J. (2007) Convolution Method and CTV-to-PTV Margins for Finite Fractions and Small Systematic Errors. Physics in Medicine & Biology, 52, 1967.
https://doi.org/10.1088/0031-9155/52/7/013

[8]   Matsuo, Y., Nakamura, M., Mizowaki, T. and Hiraoka, M. (2016) Introduction of Variance Component Analysis to Setup Error Analysis in Radiotherapy. Medical Physics, 43, 5195-5198.
https://doi.org/10.1118/1.4961397

[9]   Searle, S.R., Casella, G. and McCulloch, C.E. (2009) Variance Components (Vol. 391). John Wiley & Sons, Hoboken.

[10]   Ueda, Y., Teshima, T., Cárdenes, H. and Das, I.J. (2017) Evaluation of Initial Setup Errors of Two Immobilization Devices for Lung Stereotactic Body Radiation Therapy (SBRT). Journal of Applied Clinical Medical Physics, 18, 62-68.
https://doi.org/10.1002/acm2.12093

[11]   Cuijpers, J.P., Dahele, M., Jonker, M., Kraan, B., Senan, S., Slotman, B. and Verbakel, W.F. (2017) Analysis of Components of Variance Determining Probability of Setup Errors in CBCT-Guided Stereotactic Radiotherapy of Lung Tumors. Medical Physics, 44, 382-388.
https://doi.org/10.1002/mp.12074

[12]   Liu, H., Zhang, J., Zhong, Y., Zhang, J., Zhang, G., Deng, D, et al. (2013) The Analysis of Clinical Application of Home-Made Immobilization Device in SBRT. Chinese Journal of Radiation Oncology, 22, 501-503.

[13]   Stroom, J.C., de Boer, H.C., Huizenga, H. and Visser, A.G. (1999) Inclusion of Geometrical Uncertainties in Radiotherapy Treatment Planning by Means of Coverage Probability. International Journal of Radiation Oncology, Biology, Physics, 43, 905-919.
https://doi.org/10.1016/S0360-3016(98)00468-4

[14]   van Herk, M., Remeijer, P. and Lebesque, J.V. (2002) Inclusion of Geometric Uncertainties in Treatment Plan Evaluation. International Journal of Radiation Oncology, Biology, Physics, 52, 1407-1422.
https://doi.org/10.1016/S0360-3016(01)02805-X

[15]   Oh, S.A., Yea, J.W., Kang, M.K., Park, J.W. and Kim, S.K. (2016) Analysis of the Setup Uncertainty and Margin of the Daily Exactrac 6D Image Guide System for Patients with Brain Tumors. PloS One, 11, e0151709.
https://doi.org/10.1371/journal.pone.0151709

[16]   Herschtal, A., Foroudi, F., Silva, L., Gill, S. and Kron, T. (2012) Calculating Geometrical Margins for Hypofractionated Radiotherapy. Physics in Medicine & Biology, 58, 319.
https://doi.org/10.1088/0031-9155/58/2/319

[17]   Zimmermann, F.B., Geinitz, H., Schill, S., Thamm, R., Nieder, C., Schratzenstaller, U., et al. (2006) Stereotactic Hypofractionated Radiotherapy in Stage I (T1-2 N0 M0) Non-Small-Cell Lung Cancer (NSCLC). Acta Oncologica, 45, 796-801.
https://doi.org/10.1080/02841860600913210

[18]   Rosner, B. (1975) An Introduction to Medical Statistics. An Introduction to Medical Statistics. Oxford University Press, Oxford, 1099.
https://doi.org/10.2307/1268367

[19]   Peulen, H., Belderbos, J., Rossi, M. and Sonke, J.J. (2014) Mid-Ventilation Based PTV Margins in Stereotactic Body Radiotherapy (SBRT): A Clinical Evaluation. Radiotherapy and Oncology, 110, 511-516.
https://doi.org/10.1016/j.radonc.2014.01.010

[20]   Hinton, B., Minnich, D.J., Spencer, S.A., Popple, R. and Dobelbower, M.C. (2015) SBRT Lung: Optimal PTV Margins and Local Control. International Journal of Radiation Oncology, Biology, Physics, 93, E393.
https://doi.org/10.1016/j.ijrobp.2015.07.1549

[21]   Efron, B. (1982) The Jackknife, the Bootstrap, and Other Resampling Plans (Vol. 38). Siam.
https://doi.org/10.1137/1.9781611970319

[22]   Khan, F.M. and Gibbons, J.P. (2014) Khan’s the Physics of Radiation Therapy. Lippincott Williams & Wilkins, Philadelphia.

 
 
Top