Dosimetric Comparison of Amorphous Silicon EPID and 2D Array Detector for Pre-Treatment Verification of Intensity Modulated Radiation Therapy
Abstract: Purpose: To study the dosimetric characteristics of amorphous silicon Electronic Portal Imaging Device EPID and 2D array detector for dose verification of radiotherapy treatment plans, and the quality assurance QA testing of IMRT was investigated. Materials and methods: All measurements were done with Varian IX linear accelerator, aSi-1000 EPID and 2D array detector. The dose linearity, reproducibility, output factors, dose rate, SDD and response with slap phantom thickness have been measured and compared against those measured by ion chamber. Results: The characteristics of EPID and 2D array: the response of EPID agreed with 2D array and ion chamber 0.6cc. EPID and 2D array showed short-term output reproducibility with SD = 0.1%. The dose rates of 2D array SD = ±0.7%, EPID = ±0.4% compared with a 0.6 cc SD = ±0.5%. Output factor measurements for the central chamber of the EPID and 2D array showed no considerable deviation from ion chamber measurements. Measurement of beam profiles with the EPID and 2D array matched very well with the ion chamber measurements in the water phantom. The EPID is more sensitive to lower energy photons by increasing solid water phantom thickness. The mean and standard deviation passing rates (γ%≤1) for film, 2D array and EPID for 30 IMRT fields of five patients were 95.93 ± 0.96%, 99.05 ± 0.24%, and 99.37 ± 0.12%, respectively. Conclusion: The study shows that EPID and 2D array are a reliable and accurate dosimeter and a useful tool for quality assurance. We found that the EPID was more accurate compared with both 2D array and ion chamber. The gamma criterion of 3%/3 mm is the most suitable criteria for IMRT plans of QA.
Cite this paper: Ibrahim, A. , Mohamed, I. and Zidan, H. (2018) Dosimetric Comparison of Amorphous Silicon EPID and 2D Array Detector for Pre-Treatment Verification of Intensity Modulated Radiation Therapy. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 7, 438-452. doi: 10.4236/ijmpcero.2018.74037.
References

[1]   Grzsadziel, A., Smolińska, B., Rutkowski, R. and Slosarek, K. (2007) EPID Dosimetry-Configuration and Pre-Treatment IMRT Verification. Reports of Practical Oncology and Radiotherapy, 12, 307-312.
https://doi.org/10.1016/S1507-1367(10)60069-7

[2]   Syamkumar, S.A., Padmanabhan, S., Sukumar, P. and Nagarajan, V. (2012) Characterization of Responses of 2d Array Seven 29 Detector and Its Combined Use with Octavius Phantom for the Patient-Specific Quality Assurance in Rapidarc Treatment Delivery. Medical Dosimetry, 37, 53-60.
https://doi.org/10.1016/j.meddos.2010.12.013

[3]   Spezi, E., Angelini, A.L., Romani, F. and Ferri, A. (2005) Characterization of a 2D Ion Chamber Array for the Verification of Radiotherapy Treatments. Physics in Medicine & Biology, 50, 3361-3373.
https://doi.org/10.1088/0031-9155/50/14/012

[4]   Markovic, M., Stathakis, S., Mavroidis, P., Jurkovic, I. and Papanikolaou, N. (2014) Characterization of a Two-Dimensional Liquid-Filled Ion Chamber Detector Array Used for Verification of the Treatments in Radiotherapy. Medical Physics, 41, Article ID: 051704/14.
https://doi.org/10.1118/1.4870439

[5]   Sharma, D.S., Mhatre, V., Heigrujam, M., Talapatra, K. and Mallik, S. (2010) Portal Dosimetry for Pretreatment Verification of IMRT Plan: A Comparison with 2D Ion Chamber Array. Journal of Applied Clinical Medical Physics, 11, 238-248.
https://doi.org/10.1120/jacmp.v11i4.3268

[6]   Ezzel, G.A., Galvin, J.M., Low, D., Palta, J.R., Rosen, I. and Sharpe, M.B. (2003) Guidance on Delivery, Treatment Planning, and Clinical Implementation of IMRT: Report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Medical Physics, 30, 2089-2115.
https://doi.org/10.1118/1.1591194

[7]   Sun, W., Ren, C., Lu, N., Xu, Y., Han, F., Ping, L.Y. and Dai, J. (2016) Dosimetric Comparison of Intensity-Modulated Radiotherapy and Volumetric-Modulated Arc Radiotherapy in Patients with Prostate Cancer: A Meta-Analysis. Journal of Applied Clinical Medical Physics, 17, 254-262.
https://doi.org/10.1120/jacmp.v17i6.6464

[8]   Levendag, P.C. (2005) Dosimetric Verification of Intensity Modulated Radiotherapy with an Electronic Portal Imaging Device. Department of Radiation Oncology, Division of Medical Physics, Erasmus MC/Daniel den Hoed Cancer Center, Rotterdam.

[9]   Van Esch, A., Depuydt, T. and Huyskens, D.P. (2004) The Use of an aSi-Based EPID for Routine Absolute Dosimetric Pre-Treatment Verification of Dynamic IMRT Fields, Radiation. Oncology, 71, 223-234.
https://doi.org/10.1016/j.radonc.2004.02.018

[10]   Khan, F.M. (2003) The Physics of Radiation Therapy. 3rd Edition, Williams & Wilkins, Philadelphia.

[11]   SGSMP, SSRPM and SSRFM (2007) Quality Control for Intensity-Modulated Radiation Therapy, Recommendations No. 15.

[12]   Venencia, C.D. and Besa, P. (2004) Commissioning and Quality Assurance for Intensity Modulated Radiotherapy with Dynamic Multileaf Collimator: Experience of the Pontificia Universidad Catolica de Chle. Journal of Applied Clinical Medical Physics, 5, 37-54.

[13]   Herzen, J., Todorovic, M. and Cermers, F. (2007) Dosimetric Evaluation of a 2D Pixel Ionization Chamber for Implementation in Clinical Routine. Physics in Medicine & Biology, 52, 1197-1208.
https://doi.org/10.1088/0031-9155/52/4/023

[14]   Saminathan, S., Manickan, R., Chandraraj, V. and Supe, S.S. (2010) Dosimetric Study of 2D Ion Chamber Array Matrix for the Modern Radiotherapy Treatment Verification. Journal of Applied Clinical Medical Physics, 11, 3076.
https://doi.org/10.1120/jacmp.v11i2.3076

[15]   Peters, S., Schiefer, H. and Plasswilm, L. (2014) A Treatment Planning Study Comparing Elekta VMAT and Fixed Field IMRT Using the Varian Treatment Planning System Eclipse. Radiation Oncology, 9, 153.
https://doi.org/10.1186/1748-717X-9-153

[16]   Merheb, C., Chevillard, C., Ksouri, W., Fawzi, M., Bollet, M. and Toledano, A. (2015) Comparison between Two Different Algorithms Used for Pretreatment QA via aSi Portal Images. Journal of Applied Clinical Medical Physics, 16, 141-153.
https://doi.org/10.1120/jacmp.v16i3.5202

[17]   Boehmer, D., Bohsung, J., Eichwurzel, I., Moys, A. and Budach, V. (2004) Clinical and Physical Quality Assurance for Intensity Modulated Radiotherapy of Prostate Cancer. Radiation Oncology, 71, 319-325.
https://doi.org/10.1016/j.radonc.2004.02.016

[18]   Warkentin, B., Steciw, S., Rathee, S. and Fallone, B.G. (2003) Dosimetric IMRT Verification with a Flat-Panel EPID. Medical Physics, 30, 3143-3155.
https://doi.org/10.1118/1.1625440

[19]   Greer, P.B. and Barnes, M.P. (2007) Investigation of Amorphous Silicon EPID for Measurement and Quality Assurance of Enhanced Dynamic Wedge. Physics in Medicine and Biology, 52, 1075-1087.
https://doi.org/10.1088/0031-9155/52/4/014

[20]   Budgell, G.J., Zhang, Q., Trouncer, R.J. and Mackay, R.I. (2005) Improving IMRT Quality Control Efficiency Using an Amorphous Silicon Electronic Portal Imager. Medical Physics, 32, 3267-3278.
https://doi.org/10.1118/1.2074227

[21]   Agazaryan, N., Timothy, D.S. and Demarco, J.J. (2003) Patient Specific Quality Assurance for the Delivery of Intensity Modulated Radiotherapy. Journal of Applied Clinical Medical Physics, 4, 40-50.
https://doi.org/10.1120/jacmp.v4i1.2540

 
 
Top