 IJMPCERO  Vol.7 No.3 , August 2018
Measurement of Electron Return Effect and Skin Dose Reduction by a Bolus in an Anthropomorphic Physical Phantom under a Magnetic Resonance Guided Linear Accelerator (MR-LINAC) System
Abstract: Background: Magnetic resonance image-guided radiation therapy (MR-IGRT) promises more precise and effective radiation treatments compared to conventional IGRT by using real-time on-board MR imaging. Under the influence of a magnetic field, however, secondary electrons exiting a surface can be forced in a circular path and re-enter the medium, resulting in dose increase at a beam-exit surface, called the electron return effect (ERE). The purpose of the study is to compare the exit skin dose computed by Monte Carlo dose calculation with measurements using an adult anthropomorphic phantom and to measure the effect of skin dose reduction by adding 1 cm-thick bolus. Method: The plan was compared with measurements using an adult anthropomorphic phantom combined with radiochromic films and thermoluminescent dosimeters. We also measured the skin dose reduction by adding 1 cm-thick bolus on the frontal surface of the phantom. Results: We found that 1 cm-thick bolus reduced the skin dose by up to 20% both in measurements and calculations. The plan was found to overestimate the measured skin dose by about 10% and there was no significant difference in the bolus effect between the breast skin and the skin (without breast attachment) doses. Conclusion: In conclusion, we confirmed the ERE effect on the anthropomorphic phantom under the magnetic field and the exit skin dose reduction by adding a bolus. Skin dose measurements using anthropomorphic phantom may be helpful to evaluate more realistic skin dose and the bolus effect in the magnetic field.
Cite this paper: Han, E. , Wen, Z. , Lee, H. , Paulino, A. and Lee, C. (2018) Measurement of Electron Return Effect and Skin Dose Reduction by a Bolus in an Anthropomorphic Physical Phantom under a Magnetic Resonance Guided Linear Accelerator (MR-LINAC) System. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 7, 339-346. doi: 10.4236/ijmpcero.2018.73028.
References

[1]   Raaijmakers, A.J., et al. (2007) Dose Optimization for the MRI-Accelerator: IMRT in the Presence of a Magnetic Field. Physics in Medicine & Biology, 52, 7045-7054.
https://doi.org/10.1088/0031-9155/52/23/018

[2]   Raaijmakers, A.J., Raaymakers, B.W. and Lagendijk, J.J. (2005) Integrating a MRI Scanner with a 6 MV Radiotherapy Accelerator: Dose Increase at Tissue-Air Interfaces in a Lateral Magnetic Field Due to Returning Electrons. Physics in Medicine & Biology, 50, 1363-1376.
https://doi.org/10.1088/0031-9155/50/7/002

[3]   Raaijmakers, A.J., Raaymakers, B.W. and Lagendijk, J.J. (2007) Experimental Verification of Magnetic Field Dose Effects for the MRI-Accelerator. Physics in Medicine & Biology, 52, 4283-4291.
https://doi.org/10.1088/0031-9155/52/14/017

[4]   Raaijmakers, A.J., Raaymakers, B.W. and Lagendijk, J.J. (2008) Magnetic-Field-Induced Dose Effects in MR-Guided Radiotherapy Systems: Dependence on the Magnetic Field Strength. Physics in Medicine & Biology, 53, 909-923.
https://doi.org/10.1088/0031-9155/53/4/006

[5]   Chen, X.F., et al. (2016) Technical Note: Dose Effects of 1.5 T Transverse Magnetic Field on Tissue Interfaces in MRI-Guided Radiotherapy. Medical Physics, 43, 4797-4802.
https://doi.org/10.1118/1.4959534

[6]   Woodings, S.J., et al. (2018) Beam Characterisation of the 1.5 T MRI-Linac. Physics in Medicine & Biology, 63, 085015.
https://doi.org/10.1088/1361-6560/aab566

[7]   Oborn, B.M., et al. (2010) Monte Carlo Characterization of Skin Doses in 6 MV Transverse Field MRI-Linac Systems: Effect of Field Size, Surface Orientation, Magnetic Field Strength, and Exit Bolus. Medical Physics, 37, 5208-5217.
https://doi.org/10.1118/1.3488980

[8]   Raaymakers, B.W., et al. (2004) Integrating a MRI Scanner with a 6 MV Radiotherapy Accelerator: Dose Deposition in a Transverse Magnetic Field. Physics in Medicine & Biology, 49, 4109-4118.
https://doi.org/10.1088/0031-9155/49/17/019

[9]   Hissoiny, S., et al. (2011) GPUMCD: A New GPU-Oriented Monte Carlo Dose Calculation Platform. Medical Physics, 38, 754-764.
https://doi.org/10.1118/1.3539725

[10]   Hissoiny, S., et al. (2011) Fast Dose Calculation in Magnetic Fields with GPUMCD. Physics in Medicine & Biology, 56, 5119-5129.
https://doi.org/10.1088/0031-9155/56/16/003

[11]   Ahmad, S.B., et al. (2016) Evaluation of a Commercial MRI Linac Based Monte Carlo Dose Calculation Algorithm with GEANT4. Medical Physics, 43, 894-907.
https://doi.org/10.1118/1.4939808

[12]   ICRP (1998) Radiation Dose to Patients from Radiopharmaceuticals (Addendum 2 to ICRP Publication 53). Annals of the ICRP, 28, 1-126.

[13]   Petoussi-Henss, N., et al. (2010) ICRP Publication 116. Conversion Coefficients for Radiological Protection Quantities for External Radiation Exposures. Annals of the ICRP, 40, 1-257.
https://doi.org/10.1016/j.icrp.2011.10.001

[14]   Raaijmakers, A.J., et al. (2007) Integrating a MRI Scanner with a 6 MV Radiotherapy Accelerator: Impact of the Surface Orientation on the Entrance and Exit Dose Due to the Transverse Magnetic Field. Physics in Medicine & Biology, 52, 929-939.
https://doi.org/10.1088/0031-9155/52/4/005

 
 
Top