Back
 JAMP  Vol.6 No.5 , May 2018
Induced Electromagnetic Fields Estimation in Spine Examinations of MRI Patients: A Re-Evaluation of Existing Clinical Protocols at a Hospital in Accra, Ghana
Abstract: Patients undergoing Magnetic Resonance Imaging (MRI) are exposed to strong, non-uniform static magnetic fields outside of the central imaging region, in which the movement of the body may induce electric currents in tissues which could possibly be harmful. The purpose of this study was to re-evaluate existing clinical protocols by determining the induced electromagnetic (EM) fields in MRI spine examinations. The study covered 120 MRI spine examinations at the MRI Unit of a hospital in Accra, Ghana. A numerical model based on Faraday’s equation was developed using the finite difference method (FDM) and MATLAB software to compute, first, a test simulation of induced EM field intensities and then actual measurements of induced fields on the spine. The simulation results were peak induced electric field, 0.39 V/m and current density, 0.039 A/m2. Patient results were; calculated maximum velocity, 0.29 m/s; peak induced electric field strength, 0.44 V/m, and current density, 0.043 A/m2. The levels of induced EM-fields were such that they would not pose any potential health hazards to the patients as these values were well below the recommended guidance levels set by the Directive IEC 60601-2-33 of the European Parliament.
Cite this paper: Acheampong, F. , Dery, T. , Appiah, R. and Abaye, D. (2018) Induced Electromagnetic Fields Estimation in Spine Examinations of MRI Patients: A Re-Evaluation of Existing Clinical Protocols at a Hospital in Accra, Ghana. Journal of Applied Mathematics and Physics, 6, 1065-1075. doi: 10.4236/jamp.2018.65092.
References

[1]   Liu, F., Zhao, H. and Crozier, S. (2003) Calculation of the Electric Fields Induced by Body and Head in High-Field MRI. Journal of Magnetic Resonance, 161, 99-107.
http://pfeifer.phas.ubc.ca/refbase/files/Liu-JMagRes-2003-161-99.pdf
https://doi.org/10.1016/S1090-7807(02)00180-5


[2]   Mcrobbie, D.W. (2012) Occupational Exposure in MRI. British Journal of Radiology, 85, 293-312.
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC3486652/
https://doi.org/10.1259/bjr/30146162


[3]   ICNIRP (2014) Guidelines for Limiting Exposure to Electric Fields Induced by Movement of the Human Body in a Static Magnetic Field and by Time-Varying Magnetic Fields below 1 Hz. Health Physics, 106, 418-425.
https://www.icnirp.org/cms/upload/publications/ICNIRPmvtgdl_2014.pdf
https://doi.org/10.1097/HP.0b013e31829e5580


[4]   Directive 2004/40/EC of the European Parliament and of the Council (2004). Official Journal of the European Union, L184/1, L184-9.
https://archive.senseaboutscience.org/data/files/PhysicalAgentsDirective.pdf

[5]   Directive 2013/35/EU-Electromagnetic Fields.
https://osha.europa.eu/en/legislation/directives/directive-2013-35-eu-electromagnetic-fields

[6]   IEC 60601-2-33:2010+AMD1:2013+AMD2:2015 CSV (2010) Medical Electrical Equipment—Part 2-33: Particular Requirements for the Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnosis. Consolidated Version, 2010, 2013, 2015.
https://webstore.iec.ch/preview/info_iec60601-2-33%7Bed3.0%7Db.pdf

[7]   MR-TIP (2012) MRI.
http://www.mr-tip.org/mri

[8]   MATLAB (2013) Mathworks.
http://mathworks.com/products/matlab

[9]   Taflove, A. and Hagness, S.C. (2005) Computational Electrodynamics: The Finite-Difference Time-Domain Method. 3rd Edition, Artech House, Norwood, MA.

[10]   Green, P. and Rosenberg, J. (2010) Line Integrals around Closed Curves, and the Theorems of Green and Stokes.
https://qmplus.qmul.ac.uk/pluginfile.php/283380/mod_resource/content/3/Chapter%204%20-%20Vector%20intergration%2C%202010.pdf

[11]   Schmitt, F., Stehling, M.K. and Turner, R. (1998) Echo-Planar Imaging Theory, Technique and Application. Springer, New York, 9-14.
http://radioramanuja.com/file/download/echo-planar-imaging-theory-technique-and-application
https://doi.org/10.1007/978-3-642-80443-4


[12]   Bowtell, R. and Bowley, R.M. (2000) Analytic Calculations of the E-Fields Induced by Time-Varying Magnetic Fields Generated by Cylindrical Gradient Coils. Magnetic Resonance in Medicine, 44, 782-790.
https://doi.org/10.1002/1522-2594(200011)44:5<782::AID-MRM16>3.0.CO;2-7

 
 