Back
 IJMPCERO  Vol.7 No.1 , February 2018
Disposable Condenser Dosimeter Using a Skin-Insulated Mini-Substrate with a Silicon X-Ray Diode in Image-Guided Radiation Therapy
Abstract: To measure integral doses in image-guided radiation therapy, we developed a disposable mini-substrate with a 1.0-μF condenser and a silicon X-ray diode (Si-XD). The Si-XD is a high-sensitivity photodiode selected for detecting X-rays. In the substrate with dimensions of 15 × 15 mm2, the initial charging voltage is 3.30 V, and the charging voltage is decreased by photocurrents flowing through the Si-XD during X-ray exposing. The condenser in the substrate is charged by a microcomputer dock, and the charging voltage is also measured using an analog to digital converter in the dock after exposing X-rays. The dock is connected to a personal computer through a USB cable, and integral doses are shown on the PC monitor. The doses were proportional to decreases in the charging voltage, and the calibrated doses corresponded well to those obtained using a readily available ionization chamber.
Cite this paper: Yamaguchi, S. , Sato, E. , Nakamura, R. , Oikawa, H. , Kakuhara, H. , Kikuchi, K. , Ariga, H. and Ehara, S. (2018) Disposable Condenser Dosimeter Using a Skin-Insulated Mini-Substrate with a Silicon X-Ray Diode in Image-Guided Radiation Therapy. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 7, 35-46. doi: 10.4236/ijmpcero.2018.71004.
References

[1]   Gööket, A., Geerts, W., Hambsch, F.J., Oberstedt, S., Vidali, M. and Zeynalov, Sh. (2016) A Position-Sensitive Twin Ionization Chamber for Fission Fragment and Prompt Neutron Correlation Experiments. Nuclear Instruments and Methods in Physics Research Section A, 830, 366-374.
http://www.sciencedirect.com/science/article/pii/S0168900216305435

[2]   Gaudefroy, L., Roger, T., Pancin, J., Spitaels, C., Aupiais, J. and Mottier, J. (2017) A Twin Frisch-Grid Ionization Chamber as a Selective Detector for the Delayed Gamma-Spectroscopy of fission Fragments. Nuclear Instruments and Methods in Physics Research Section A, 855, 133-139.
http://www.sciencedirect.com/science/article/pii/S0168900217302711

[3]   Cederlund, J. (1960) A Three-Electrode Ionization Condenser Chamber. British Journal of Radiology, 33, 105-109.
https://doi.org/10.1259/0007-1285-33-386-105

[4]   Matsushita, R., Sato, E. , Yanbe, Y., Chiba, H., Maeda, T., Hagiwara, O., Matsukiyo, H., Osawa, A., Enomoto, T., Watanabe, M., Kusachi, S., Sato, S., Ogawa, A. and Onagawa, J. (2013) Low-Dose-Rate Computed Tomography System Utilizing 25 mm/s-Scan Silicon X-Ray Diode and Its Application to Iodine K-Edge Imaging Using Filtered Bremsstrahlung Photons. Japanese Journal of Applied Physics, 52, 032202-1-5.
http://iopscience.iop.org/article/10.7567/JJAP.52.032202
https://doi.org/10.7567/JJAP.52.032202


[5]   Sato, Y., Sato, E., Ehara, S., Oda, Y., Hagiwara, O., Matsukiyo, H., Enomoto, T., Watanabe, M. and Kusachi, S. (2015) Development of a Dual-Energy Silicon X-Ray Diode and Its Application to Gadolinium Imaging. Radiation Measurement, 77, 12-17.
http://www.sciencedirect.com/science/article/pii/S1350448715300044
https://doi.org/10.1016/j.radmeas.2015.04.003


[6]   Groh, B.A., Siewerdsen, J.H., Drake, D.G., Wong, J.W. and Jaffray, D.A. (2002) A Performance Comparison of Flat-Panel Imager-Based MV and kV Cone-Beam CT. Medical Physics, 29, 967-975.
https://doi.org/10.1118/1.1477234

[7]   Thapa, B.B. and Molloy, J.A. (2013) Feasibility of an Image Planning System for Kilovoltage Image-Guided Radiation Therapy. Medical Physics, 40, 061703-1-10.
https://doi.org/10.1118/1.4803508

[8]   Kanamoto, M., Miyati, T., Terashima, K., Suga, D. and Fuwa, N. (2016) Preliminary Study of Apparent Diffusion Coefficient Assessment after Ion Beam Therapy for Hepatocellular Carcinoma. Radiological Physics and Technology, 9, 233-239.
https://link.springer.com/article/10.1007/s12194-016-0354-5

[9]   Onizuka, R., Araki, F., Ohno, T., Nakaguchi, Y., Kai, Y., Tomiyama, Y. and Hioki, K. (2016) Accuracy of Dose Calculation Algorithms for Virtual Heterogeneous Phantoms and Intensity-Modulated Radiation Therapy in the Head and Neck. Radiological Physics and Technology, 9, 77-87.
https://link.springer.com/article/10.1007%2Fs12194-015-0336-z
https://doi.org/10.1007/s12194-015-0336-z

[10]   Takakura, T., Ito, Y., Higashikawa, A., Nishiyama, T. and Sakamoto, T. (2016) Verification of the Dose Attenuation of a Newly Developed Vacuum Cushion for Intensity-Modulated Radiation Therapy of Prostate Cancer. Radiological Physics and Technology, 9, 270-276.
https://link.springer.com/article/10.1007%2Fs12194-016-0359-0
https://doi.org/10.1007/s12194-016-0359-0

[11]   Arakawa, Y., Sato, E., Kogita, H., Hamaya, T., Nihei, S., Numahata, W., Kami, S., Oda, Y., Hagiwara, O., Matsukiyo, H., Osawa, A., Enomoto, T., Watanabe, M., Kusachi, S., Sato, S. and Ogawa, A. (2014) Investigation of X-Ray Photon-Counting Using Ceramic-Substrate Silicon Diode and Its Application to Gadolinium Imaging. Japanese Journal of Applied Physics, 53, 072201-1-5.
http://iopscience.iop.org/article/10.7567/JJAP.53.072201
https://doi.org/10.7567/JJAP.53.072201


[12]   (2017) Measure Works.
https://www.measureworks.co.jp/Directreadingdosimeter.htm

 
 
Top