TBI Technique Improvements for Anesthetized Pediatric Patients Based on Near-Miss Incident Reporting
Abstract: Purpose: To share our clinical experience of an optimized and comprehensive pediatric TBI technique. Methods and Materials: Through the use of incident learning, safety-critical areas were identified in our procedure for total body irradiation (TBI) for pediatric patients under anesthesia for bone-marrow transplant. The previous procedure lacked flexibility to accommodate various requests from the anesthesia team due to the wide range of patient sizes. To address this issue and to improve the process overall, we updated our procedure for TBI simulation, dosimetry planning, patient setup during treatment, and in vivo dosimetry. A simulation form was redesigned with additional detailed instructions and documentation requirements. The dose calculation procedure was reformulated to remove dependence on setup variations. Tissue compensation determination and therefore dose uniformity were improved by introducing rigorous calculation methods. Calculations were performed on 28 previously-treated patients to compare the dose uniformity using the new versus previous methods. Results: The new procedures improve interdepartmental communication, simplify the workflow, decrease the risk of treating patients in a setup that differs from that used during the simulation, and reduce dose heterogeneity. The new compensator design significantly improved patient dose uniformity: 0.8% ± 0.4% (new method) vs. 4.2% ± 2.3% (previous method) (p < 0.01). Conclusion: A near-miss incident reporting system was used to improve the safety and quality of pediatric TBI procedures under anesthesia.
Cite this paper: Cao, N. , Ford, E. , Nyflot, M. , Ermoian, R. , Pelly, N. , Young, L. and Hendrickson, K. (2016) TBI Technique Improvements for Anesthetized Pediatric Patients Based on Near-Miss Incident Reporting. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 5, 270-280. doi: 10.4236/ijmpcero.2016.54027.
References

[1]   Pasquini, M.C. and Wang, Z. (2009) Current Use and Outcome of Hematopoietic Stem Cell Transplantation: Part I—Cibmtr Summary Slides. CIBMTR Newsletter [Serial Online], 15, 7-11.

[2]   Bunin, N., Aplenc, R., Kamani, N., Shaw, K., Cnaan, A. and Simms, S. (2003) Randomized Trial of Busulfan vs Total Body Irradiation Containing Conditioning Regimens for Children with Acute Lymphoblastic Leukemia: A Pediatric Blood and Marrow Transplant Consortium Study. Bone Marrow Transplantation, 32, 543-548.
http://dx.doi.org/10.1038/sj.bmt.1704198

[3]   Davies, S.M., Ramsay, N.K., Klein, J.P., Weisdorf, D.J., Bolwell, B., Cahn, J.Y., Camitta, B,M., Gale, R.P., Giralt, S., Heilmann, C., Henslee-Downey, P.J., Herzig, R.H., Hutchinson, R., Keating, A., Lazarus, H.M., Milone, G.A., Neudorf, S., Perez, W.S., Powles, R.L., Prentice, H.G., Schiller, G., Socié, G., Vowels, M., Wiley, J., Yeager, A. and Horowitz, M.M. (2000) Comparison of Preparative Regimens in Transplants for Children with Acute Lymphoblastic Leukemia. Journal of Clinical Oncology, 18, 340-347.

[4]   Aschan, J. (2007) Risk Assessment in Haematopoietic Stem Cell Transplantation: Conditioning. Best Practice & Research Clinical Haematology, 20, 295-310.
http://dx.doi.org/10.1016/j.beha.2006.09.004

[5]   Dyk, L.V., Galvin, J.M., Glasgow, G.P. and Podgorsak, E.B. (1986) The Physical Aspects of Total and Half Body Photon Irradiation. AAPM Report No. 17 (Task Group 29), American Institute of Physics, College Park.

[6]   Halperin, E.C., Brady, L.W., Perez, C.A. and Wazer, D.E. (2013) Perez and Brady’s Principles and Practice of Radiation Oncology. 6th Edition, Lippincott Williams & Wilkins, Philadephia.

[7]   Yukihara, E.G., Mardirossian, G., Mirzasadeghi, M., Guduru, S. and Ahmad, S. (2008) Evaluation of Al2O3: C optically Stimulated Luminescence (OSL) Dosimeters for Passive Dosimetry of High-Energy Photon and Electron Beams in Radiotherapy. Medical Physics, 35, 260-269.
http://dx.doi.org/10.1118/1.2816106

[8]   Huq, M.S., Fraass, B.A., Dunscombe, P.B., Gibbons Jr., J.P., Ibbott, G.S., Mundt, A.J., Mutic, S., Palta, J.R., Rath, F., Thomadsen, B.R., Williamson, J.F. and Yorke, E.D. (2016) The Report of Task Group 100 of the AAPM: Application of Risk Analysis Methods to Radiation Therapy Quality Management. Medical Physics, 43, 4209.
http://dx.doi.org/10.1118/1.4947547

[9]   Vollans, S.E., Perrin, B., Wilkinson, J.M., Gattamaneni, H.R. and Deakin, D.P. (2000) Investigation of Dose Homogeneity in Paediatric Anthropomorphic Phantoms for a Simple Total Body Irradiation Technique. The British Journal of Radiology, 73, 317-321.
http://dx.doi.org/10.1259/bjr.73.867.10817050

[10]   Wills, C., Cherian, S., Jacob Yousef, J., Wang, K. and Mackley, H.B. (2016) Total Body Irradiation: A Practical Review. Applied Radiation Oncology, 5, 11-17.

[11]   Mesa, F., Eng, T.Y., Esquivel, C., Fuller, C.D., Papanikolaou, N. and Sosa, M. (2011) Implementation of a Lateral Total Body Irradiation Technique with 6 MV Photons: The University of Texas Health Science Center in San Antonio Experience. Journal of Radiotherapy in Practice, 10, 45-54.
http://dx.doi.org/10.1017/S1460396910000221

[12]   Hui, S.K., Kapatoes, J., Fowler, J., Henderson, D., Olivera, G., Manon, R.R., Gerbi, B., Mackie, T.R. and Welsh, J.S. (2005) Feasibility Study of Helical Tomotherapy for Total Body or Total Marrow Irradiation. Medical Physics, 32, 3214-3224.
http://dx.doi.org/10.1118/1.2044428

[13]   Gruen, A., Ebell, W., Wlodarczyk, W., Neumann, O., Kuehl, J.S., Stromberger, C., Budach, V. and Marnitz, S. (2013) Total Body Irradiation (TBI) Using Helical Tomotherapy in Children and Young Adults Undergoing Stem Cell Transplantation. Radiation Oncology, 8, 92.
http://dx.doi.org/10.1186/1748-717X-8-92

 
 
Top