OJAPr  Vol.3 No.4 , December 2015
Study of Correlation Coefficient for Breast Tumor Detection in Microwave Tomography
ABSTRACT
In microwave tomography, it is necessary to increase the amount of diverse observation data for accurate image reconstruction of the dielectric properties of the imaging area. The multi-polarization method has been proposed as a suitable technique for the acquisition of a variety of observation data. While the effectiveness of employing multi-polarization to reconstruct images has been confirmed, the physical considerations related to image reconstruction have not been investigated. In this paper, a compact-sized imaging sensor using multi-polarization for breast cancer detection is presented. An analysis of the correlation coefficient of the received data of adjacent antennas was performed to interpret the imaging results. Numerical simulation results demonstrated that multi-polarization can reconstruct images better compared to single polarizations owing to its low correlation coefficient and condition number.

Cite this paper
Mohamed, L. , Ozawa, N. , Ono, Y. , Kamiya, T. and Kuwahara, Y. (2015) Study of Correlation Coefficient for Breast Tumor Detection in Microwave Tomography. Open Journal of Antennas and Propagation, 3, 27-36. doi: 10.4236/ojapr.2015.34004.
References
[1]   Saika, K. and Sobue, T. (2009) Epidemiology of Breast Cancer in Japan and the US. Japan Medical Association Journal, 52, 39-44.

[2]   Nikolova, N.K. (2011) Microwave Imaging for Breast Cancer. IEEE Microwave Magazine, 12, 78-94. http://dx.doi.org/10.1109/MMM.2011.942702

[3]   Lazebnik, M., Popovic, D., McCartney, L., Watkins, C.B., Lindstrom, M.J., Harter, J., Sewall, S., Ogilvie, T., Magliocco, A., Breslin, T.M., Temple, W., Mew, D., Booske, J.H., Okonniewsk, M. and Hagness, S.C. (2007) A Large Scale Study of the Ultra-Wideband Microwave Dielectric Properties of Normal, Benign, and Malignant Breast Tissues Obtained from Cancer Surgeries. Physics in Medicine and Biology, 52, 6093-6115. http://dx.doi.org/10.1088/0031-9155/52/20/002

[4]   Kuwahara, Y., Miura, S., Nishina, Y., Mukumoto, K., Ogura, H. and Sakahara, H. (2013) Clinical Setup of Microwave Mammography. IEICE Transactions Communication, E96-B, 2553-2562. http://dx.doi.org/10.1587/transcom.E96.B.2553

[5]   Kuwahara, Y., Ogura, H. and Sakahara, H. (2013) Microwave Mammography—Considerations on Clinical Test. MWE 2013 Microwave Workshop Digest, 50-53.

[6]   Kuwahara, Y., Ogura, H. and Sakahara, H. (2013) Microwave Mammography—Considerations on Clinical Test. Proceedings of the 26th Annual International Conference of the IEEE EMBS, San Francisco, 1-5 September 2004, 2758- 2761.

[7]   Kuwahara, Y. (2013) Microwave Mammography Technology (Japanese). RF World, 25, 8-19.

[8]   Ozawa, N. and Kuwahara, Y. (2014) Considerations of Antennas for Microwave Mammography. Proceedings of Thailand-Japan Microwave (TJMW), TH3-4.

[9]   Mohamed, L. and Kuwahara, Y. (2014) Distortion Born Iterative Method in Microwave Tomography—A Numerical Study of 3D Non-Debye and Debye Model. Proceedings of Thailand-Japan Microwave (TJMW), TH4-1.

[10]   Meaney, P.M., Fanning, M.W., Li, D., Poplack, S.P. and Paulsen, K.D. (2000) A Clinical Prototype for Active Microwave Imaging of the Breast. IEEE Transactions on Microwave Theory and Techniques, 48, 1841-1853. http://dx.doi.org/10.1109/22.883861

[11]   Fang, Q., Meaney, P.M. and Paulsen, K.D. (2004) Microwave Image Reconstruction of Tissue Property Dispersion Characteristics Utilizing Multiple Frequency Information. IEEE Transactions on Microwave Theory and Techniques, 52, 1866-1875. http://dx.doi.org/10.1109/TMTT.2004.832014

[12]   Ostadrahimi, M., Zakaria, A., LoVetri, J. and Shafai, L. (2013) A Near-Field Dual Polarized (TE-TM) Microwave Imaging System. IEEE Transactions on Microwave Theory and Techniques, 61, 1376-1384. http://dx.doi.org/10.1109/TMTT.2012.2237181

[13]   Mays, R.O., Behdad, N. and Hagness, S.C. (2015) A TSVD Analysis of the Impact of Polarization on Microwave Breast Imaging Using an Enclosed Array of Miniaturized Patch Antennas. IEEE Antennas and Wireless Propagation Letters, 14, 418-421. http://dx.doi.org/10.1109/LAWP.2014.2365755

[14]   Joachimowicz, N., Pichot, C. and Hugonin, J.P. (1991) Inverse Scattering: An Iterative Numerical Method for Electromagnetic Imaging. IEEE Transactions on Antennas and Propagation, 39, 1742-1752. http://dx.doi.org/10.1109/8.121595

[15]   Shea, J.D., Kosmas, P., Veen, V. and Hagness, S.C. (2010) Contrast-Enhanced Microwave Imaging of Breast Tumors: A Computational Study Using 3D Realistic Numerical Phantoms. Inverse Problem, 26, 1-22. http://dx.doi.org/10.1088/0266-5611/26/7/074009

[16]   Shea, J.D., Kosmas, P., Hagness, S.C. and Veen, B.D.V. (2010) Three-Dimensional Microwave Imaging of Realistic Numerical Breast Phantoms via a Multiple Frequency Inverse Scattering Technique. Medical Physics, 37, 4210-4226. http://dx.doi.org/10.1118/1.3443569

[17]   Semenov, S., Kellam, J., Sizov, Y., Nazarov, A., Williams, T., Nair, B., Pavlovsky, A., Posukh, V. and Quinn, M. (2011) Microwave Tomography of Extremities: 1. Dedicated 2-D System and Physiological Signatures. Physics in Medicine and Biology, 56, 2005-2017. http://dx.doi.org/10.1088/0031-9155/56/7/006