IJMPCERO  Vol.4 No.4 , November 2015
The Measurement Accuracy of Ball Bearing Center in Portal Images Using an Intensity-Weighted Centroid Method
ABSTRACT
Medical linac based imaging modalities such as portal imaging can be utilized for highly accurate measurements. An intensity-weighted centroid method for determining object center is proposed that can detect the position of small object at subpixel accuracy. The principles and algorithms of the intensity-weighted centroid method are presented. Analytical results are derived for positional accuracy of a rod and a sphere in digital images, and the theoretical accuracy limits are calculated. The method was experimentally examined using phantoms with embedded ball bearings (BBs). Images of the phantoms were taken by the MV portal imager of a medical linac. The image pixel size was 0.26 mm when projected at the linac isocenter plane. The BB coordinates were calculated by applying the intensity-weighted centroid method after removing the background. The reproducibility of BB position detection was measured with 3 monitor unit (MU) exposures at various dose rates. A stationary BB, of 0.25 image contrast, showed position reproducibility in the range of 0.004 - 0.013 mm. When the method was used to measure the displacement of a moving BB, the difference between the measured and expected BB position had a standard deviation of 0.006 mm. The effect of image noise on the BB detection accuracy was measured using a phantom with multiple BBs. The overall detection accuracy, represented by standard deviation, steadily improved from 0.13 mm at 0.03 MU to 0.008 mm at 5.0 MU, and showed an inverse correlation with contrast-to-noise ratio. We demonstrated that intensity-weighted centroid method can achieve subpixel accuracy in position detection. With a linac based imaging system, precise mechanical measurement with accuracy of microns could be achieved.

Cite this paper
Zhang, M. , Driewer, J. , Zhang, Y. , Zhou, S. and Zhu, X. (2015) The Measurement Accuracy of Ball Bearing Center in Portal Images Using an Intensity-Weighted Centroid Method. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 273-283. doi: 10.4236/ijmpcero.2015.44033.
References
[1]   Klein, E.E., Hanley, J., Bayouth, J., Yin, F.F., Simon, W., Dresser, S., et al. (2009) Task Group 142 Report: Quality Assurance of Medical Accelerators. Medical Physics, 36, 4197-4212.
http://dx.doi.org/10.1118/1.3190392

[2]   Lutz, W., Winston, K.R. and Maleki, N. (1988) A System for Stereotactic Radiosurgery with a Linear Accelerator. International Journal of Radiation Oncology*Biology*Physics, 14, 373-381.
http://dx.doi.org/10.1016/0360-3016(88)90446-4

[3]   Sykes, J.R., Lindsay, R., Dean, C.J., Brettle, D.S., Magee, D.R. and Thwaites, D.I. (2008) Measurement of Cone Beam CT Coincidence with Megavoltage Isocentre and Image Sharpness Using the QUASAR Penta-Guide Phantom. Physics in Medicine and Biology, 53, 5275-5293.
http://dx.doi.org/10.1088/0031-9155/53/19/002

[4]   Du, W., Yang, J., Luo, D. and Martel, M. (2010) A Simple Method to Quantify the Coincidence between Portal Image Graticules and Radiation Field Centers or Radiation Isocenter. Medical Physics, 37, 2256-2263.
http://dx.doi.org/10.1118/1.3397452

[5]   Du, W., Yang, J.N., Chang, E.L., Luo, D., McAleer, M.F., Shiu, A. and Martel, M.K. (2010) A Quality Assurance Procedure to Evaluate Cone-Beam CT Image Center Congruence with the Radiation Isocenter of a Linear Accelerator. Journal of Applied Clinical Medical Physics, 11, 15-26.

[6]   Grimm, J., Grimm, S.L., Das, I.J., Zhu, Y., Yeo, I., Xue, J., et al. (2011) A Quality Assurance Method with Submillimeter Accuracy for Sterotactic Linear Accelerators. Journal of Applied Clinical Medical Physics, 12, 182-198.

[7]   Du, W. and Gao, S. (2011) Measuring the Wobble of Radiation Field Centers during Gantry Rotation and Collimator Movement on a Linear Accelerator. Medical Physics, 38, 4575-4578.
http://dx.doi.org/10.1118/1.3609098

[8]   Nyflot, M.J., Cao, N., Meyer, J. and Ford, E.C. (2014) Improved Accuracy for Noncoplanar Radiotherapy: An EPID-Based Method for Submillimeter Alignment of Linear Accelerator Table Rotation with MV Isocenter. Journal of Applied Clinical Medical Physics, 15, 151-159.

[9]   Gao, S., Du, W., Balter, P., Munro, P. and Jeung, A. (2014) Evaluation of IsoCal Geometric Calibration System for Varian Linacs Equipped with On-Board Imager and Electronic Portal Imaging Device Imaging Systems. Journal of Applied Clinical Medical Physics, 15, 164-181.

[10]   Schreibmann, E., Elder, E. and Fox, T. (2009) Automated Quality Assurance for Image-Guided Radiation Therapy. Journal of Applied Clinical Medical Physics, 10, 71-79.
http://dx.doi.org/10.1120/jacmp.v10i1.2919

[11]   Winey, B., Sharp, G. and Bussière, M. (2011) A Fast Double Template Convolution Isocenter Evaluation Algorithm with Subpixel Accuracy. Medical Physics, 38, 223-227.
http://dx.doi.org/10.1118/1.3524227

[12]   Dong, L., Shiu, A., Tung, S. and Boyer, A. (1997) Verification of Radiosurgery Target Point Alignment with an Electronic Portal Imaging Device (EPID). Medical Physics, 24, 263-267.
http://dx.doi.org/10.1118/1.598070

[13]   Zhang, M., Huang, M., Le, C., Zanzonico, P.B., Claus, F., Kolbert, K.S., et al. (2008) Accuracy and Reproducibility of Tumor Positioning during Prolonged and Multi-Modality Animal Imaging Studies. Physics in Medicine and Biology, 53, 5867-5882.
http://dx.doi.org/10.1088/0031-9155/53/20/021

[14]   Du, W. and Yang J. (2009) A Robust Hough Transform Algorithm for Determining the Radiation Centers of Circular and Rectangular Fields with Subpixel Accuracy. Physics in Medicine and Biology, 54, 555-567.
http://dx.doi.org/10.1088/0031-9155/54/3/006

[15]   Sindelar, C.V. and Grigorieff, N. (2011) An Adaptation of the Wiener Filter Suitable for Analyzing Images of Isolated Single Particles. Journal of Structural Biology, 176, 60-74.
http://dx.doi.org/10.1016/j.jsb.2011.06.010

[16]   Otsu, N. (1979) A Threshold Selection Method from Gray-Level Histograms. IEEE Transactions on Systems, Man, and Cybernetics, 9, 62-66.
http://dx.doi.org/10.1109/TSMC.1979.4310076

[17]   Zhang, M., Zhou S. and Qu T. (2015) What Do We Mean When We Talk about the Linac Isocenter? International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 233-242.
http://dx.doi.org/10.4236/ijmpcero.2015.43028

 
 
Top