Measurement of Transcutaneous Bilirubin with Bilicheck as a Jaundice Screening Method in Neonates in Pediatric Emergency Departments
ABSTRACT
Objectives: To study the reliability of a transcutaneous bilirubinometer (Bilicheck) to determine bilirubin levels in neonates consulting for jaundice in a Paediatric Emergency Department (ED), and to evaluate its usefulness as a screening method. Methods: Prospective observational study realized between June of 2005 and December of2005 inneonates consulting at a paediatric emergency department for jaundice, in whom we realized both transcutaneous and total serum bilirubin measurements (TcB and TSB). We collected demographic variables, analytical variables (serum and transcutaneous bilirubin levels), length of stay in the ED, and need for treatment. Results: 66 children were included aged 2 to 31 days (81% of the sample were 2 to 7 days old). There was a close and statistically significant correlation between TcB and TSB (r = 0.81, p < 0.001). The area under the ROC curve was of 0.90, allowing detecting newborns with jaundice susceptible of treatment with TcB levels ≥ 13 mg/dL (sensitivity 92%, specificity 63, 5%, a positive predictive value 39% and a negative predictive value 97%). The number of venous punctures could be reduced in 50%. The medium stay in the ED was of 2 hours when performing serum measurements. Conclusions: A linear correlation exists between TcB-TSB. TcB measurement cannot replace that of TsB, however it could be used as a screening method in an ED to determine which neonates need confirmation by TsB measurement. The use of transcutaneous bilirubinometer would reduce both the number of painful interventions in neonates and the medium length of stay in ED, consequently reducing iatrogenesis.

Cite this paper
Míguez, C. , Salto, M. and Marañón, R. (2015) Measurement of Transcutaneous Bilirubin with Bilicheck as a Jaundice Screening Method in Neonates in Pediatric Emergency Departments. Open Journal of Pediatrics, 5, 240-247. doi: 10.4236/ojped.2015.53037.
References
[1]   American Academy of Pediatrics, Subcommittee on Hyperbilirubinemia (2004) Management of Hyperbilirubinemia in the Newborn Infant 35 or More Weeks of Gestation [Published Correction Appears in Pediatrics. 2004, 114, 1138]. Pediatrics, 114, 297-316.
http://dx.doi.org/10.1542/peds.114.1.297

[2]   Canadian Paediatric Society, Fetus and Newborn Committee (2007) Guidelines for Detection, Management and Prevention of Hyperbilirubinemia in Term and Late Preterm Newborn Infants (35 or More Weeks’ Gestation). Paediatr Child Health, 12, 401-407.

[3]   Maisels, M.J., Bhutani, V.K., Bogen, D., et al. (2009) Hyperbilirubinemia in the Newborn Infant 9 or = 35 Weeks’ Gestation: An Update with Clarifications. Pediatrics, 124, 1193-1198.
http://dx.doi.org/10.1542/peds.2009-0329

[4]   Keren, R., Tremont, K., Luan, X. and Cnaan, A. (2009) Visual Assessment of Jaundice in Term and Late Preterm Infants. Archives of Disease in Childhood—Fetal and Neonatal Edition, 94, F317-F322.
http://dx.doi.org/10.1136/adc.2008.150714

[5]   Kaplan, M., Shchors, I., Algur, N., et al. (2008) Visual Screening versus Transcutaneous Bilirubinometry for Predischarge Jaundice Assessment. Acta Paediatrica, 97, 759-763.
http://dx.doi.org/10.1111/j.1651-2227.2008.00807.x

[6]   Moyer, V.A., Ahn, C. and Sneed, S. (2000) Accuracy of Clinical Judgment in Neonatal Jaundice. Archives of Pediatrics and Adolescent Medicine, 154, 391-394.

[7]   Samanta, S., Tan, M., Kissack, C., et al. (2004) The Value of Bilicheck as Screening Tool for Neonatal Jaundice in Term and Near-Term Babies. Acta Paediatrica, 93, 1486-1490.
http://dx.doi.org/10.1111/j.1651-2227.2004.tb02634.x

[8]   Nanjundaswamy, S., Petrova, A., Mehta, R., Bernstein, W. and Hegyi, T. (2004) The Accuracy of Transcutaneous Bilirubin Measurements in Neonates: A Correlation Study. Biology of the Neonate, 85, 21-25.
http://dx.doi.org/10.1159/000074953

[9]   Wong, C.M., van Dijk, P.J.E. and Laing, I.A. (2002) A Comparison of Transcutaneous Bilirubinometers: SpectRx BiliCheck versus Minolta AirShields. Archives of Disease in Childhood—Fetal and Neonatal Edition, 87, 137-140.
http://dx.doi.org/10.1136/fn.87.2.F137

[10]   Bhutani, V.K., Gourley, G.R., Adler, S., Kreamer, B., Dalin, C. and Johnson, L.H. (2000) Noninvasive Measurement of Total Serum Bilirubin in a Multiracial Predischarge Newborn Population to Assess the Risk of Severe Hyperbilirubinemia. Pediatrics, 106, E17.
http://dx.doi.org/10.1542/peds.106.2.e17

[11]   Maisels, M.J. (2006) Transcutaneous Bilirubinometry. Neoreviews, 7, e217-e225.
http://dx.doi.org/10.1542/neo.7-5-e217

[12]   Samanta, S., Tan, M., Kissack, C., Nayak, S., et al. (2004) The Value of Bilicheck as a Screening Tool for Neonatal Jaundice in Term an Near-Term Babies. Acta Paediatrica, 93, 1486-1490.
http://dx.doi.org/10.1111/j.1651-2227.2004.tb02634.x

[13]   Rubaltelli, F., Gourley, G., Loskamp, N., et al. (2001) Transcutaneous Bilirrbin Measurement a Multicenter Evaluation of a New Device. Pediatrics, 107, 1264-1271.
http://dx.doi.org/10.1542/peds.107.6.1264

[14]   Thayyil, S. and Marriott, L. (2005) Can Transcutaneous Bilirubinometry Reduce the Need for Serum Bilirubin Estimations in Term and near Term Infants? Archives of Disease in Childhood, 90, 1311-1312.
http://dx.doi.org/10.1136/adc.2004.070292

[15]   Bertini, G. and Rubaltelli, F. (2002) Non-Invasive Bilirubinometry in Neonatal Jaundice. Seminars in Neonatology, 7, 129-133.
http://dx.doi.org/10.1053/siny.2002.0100

[16]   da Cunha Leite, M. das G., de Araújo Granato, V., Facchini, F.P. and Marba, S.T.M. (2007) Comparison of Transcutaneous and Plasma Bilirubin Measurement. Jornal de Pediatria, 83, 283-286.
http://dx.doi.org/10.2223/JPED.1619

[17]   Fouzas, S., Karatza, A.A., Skylogianni, E., Mantagou, L. and Varvarigou, A. (2010) Transcutaneous Bilirubin Levels in Late Preterm Neonates. Journal of Pediatrics, 157, 762-766.
http://dx.doi.org/10.1016/j.jpeds.2010.04.076

[18]   Zecca, E., Barone, G., De Luca, D., Marra, R., Tiberi, E. and Romagnoli, C. (2009) Skin Bilirubin Measurement during Phototherapy in Preterm and Term Newborn Infants. Early Human Development, 85, 537-540.
http://dx.doi.org/10.1016/j.earlhumdev.2009.05.010

[19]   Yu, Z.B., Dong, X.Y., Han, S.P., Chen, Y.L., Qiu, Y.F., Sha, L., et al. (2011) Transcutaneous Bilirubin Nomogram for Predicting Neonatal Hyperbilirubinemia in Healthy Term and Late-Preterm Chinese Infants. European Journal of Pediatrics, 170, 185-191.
http://dx.doi.org/10.1007/s00431-010-1281-9

[20]   Mishra, S., Chawla, D., Agarwal, R., Deorari, A.K. and Paul, V.K. (2010) Transcutaneous Bilirubin Levels in Healthy Term and Late Preterm Indian Neonates. Indian Journal of Pediatrics, 77, 45-50.
http://dx.doi.org/10.1007/s12098-010-0007-3

[21]   Wainer, S., Rabi, Y., Parmar, S.M., Allegro, D. and Lyon, M. (2009) Impact of Skin Tone on the Performance of a Transcutaneous Jaundice Meter. Acta Paediatrica, 98, 1909-1915.
http://dx.doi.org/10.1111/j.1651-2227.2009.01497.x

[22]   De Luca, D., Jackson, G.L., Tridente, A., Carnielli, V.P. and Engle, W.D. (2009) Transcutaneous Bilirubin Nomograms: A Systematic Review of Population Differences and Analysis of Bilirubin Kinetics. Archives of Pediatrics & Adolescent Medicine, 163, 1054-1059.
http://dx.doi.org/10.1001/archpediatrics.2009.187

[23]   Lam, T.S., Tsui, K.L. and Kam, C.W. (2008) Evaluation of a Point of Care Transcutaneous Bilirubinometer in Chinese Neonates at an Accident and Emergency Department. Hong Kong Medical Journal, 14, 356-360.

[24]   Yamamoto, L.G., Killeen, J. and French, G.M. (2012) Transcutaneous Bilirubin Measurement Methods in Neonates and Its Utility for Emergency Department Use. Pediatric Emergency Care, 28, 380-387.
http://dx.doi.org/10.1097/PEC.0b013e31824dcb43

 
 
Top