IJMPCERO  Vol.4 No.3 , August 2015
Evaluation of Mid-Plane Dose by Flat Panel Transit Dosimetry Method and by EBT2 Film Incident Dosimetry
ABSTRACT
Purpose: The aim of this study is to validate an easily applicable flat panel dosimetry method based on the back projection approach and to compare this method with the incident dosimetry by EBT2 film method for mid-plane dose calculations. Methods: The dosimetric characteristics of the flat panel were determined for 6 MV photon energy. Then, the methodology to calculate the dose on the central axis of the photon beam was described. While, the flat panel dosimetry method was validated with phantom measurements using an ionization chamber. Once the method was validated, in vivo measurements of ten prostate patients treated with 6 MV photon energy 3D conformal plans were also performed both with the flat panel and the EBT2 films. Results: The phantom measurements revealed a mean dispersion of 1.7% between flat panel and ionization chamber doses and 2.2% between flat panel and EBT2 film doses. While, the in vivo measurements in prostate patients revealed a mean dispersion of 0.8% between flat panel doses and treatment planning calculated doses and 1% between flat panel and EBT2 film doses. Conclusions: The presented flat panel dosimetry method is accurate, easily applicable to all types of flat panels without the use of any sophisticated software and is not time consuming.

Cite this paper
Mandourari, K. , Antypas, C. , Armpilia, C. , Floros, I. , Karaiskos, P. and Sandilos, P. (2015) Evaluation of Mid-Plane Dose by Flat Panel Transit Dosimetry Method and by EBT2 Film Incident Dosimetry. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 243-254. doi: 10.4236/ijmpcero.2015.43029.
References
[1]   Wendling, M., Louwe, R.J.W., McDermott, L.N., Sonke, J.J., Van Herk, M. and Mijnheer, B.J. (2006) Accurate Two-Dimensional IMRT Verification Using a Back-Projection EPID Dosimetry Method. Medical Physics, 33, 259-273.
http://dx.doi.org/10.1118/1.2147744

[2]   International Atomic Energy Agency (2013) Development of Procedures for in Vivo Dosimetry in Radiotherapy. IAEA Human Health Report No. 8.

[3]   Nijsten, S.M.J.J.G., Van Elmpt, W.J.C., Jacobs, M., Mijnheer, B.J., Dekker, A.L.A.J., Lambin, P., et al. (2007) A Global Calibration Model for a-Si EPIDs Used for Transit Dosimetry. Medical Physics, 34, 3872-3884.
http://dx.doi.org/10.1118/1.2776244

[4]   McDermott, L.N., Louwe, R.J.W., Sonke, J.J., Van Herk, M.B. and Mijnheer, B.J. (2004) Dose-Response and Ghosting Effects of an Amorphous Silicon Electronic Portal Imaging Device. Medical Physics, 31, 285-295.
http://dx.doi.org/10.1118/1.1637969

[5]   McDermott, L.N., Wendling, M., Sonke, J.J., Van Herk, M. and Munheer, B.J. (2007) Replacing Pretreatment Verification with in Vivo EPID Dosimetry for Prostate IMRT. International Journal of Radiation Oncology*Biology*Physics, 67, 1568-1577.
http://dx.doi.org/10.1016/j.ijrobp.2006.11.047

[6]   Herman, M.G., Balter, J.M., Jaffray, D.A., McGee, K.P., Munro, P., Shalev, S., et al. (2001) Clinical Use of Electronic Portal Imaging: Report of AAPM Radiation Therapy Committee Task Group 58. Medical Physics, 28, 712-737.
http://dx.doi.org/10.1118/1.1368128

[7]   Menon, G.V. and Sloboda, R.S. (2003) Compensator Quality Control with an Amorphous Silicon EPID. Medical Physics, 30, 1816-1824.
http://dx.doi.org/10.1118/1.1584040

[8]   Chang, J., Mageras, G.S., Ling, C.C. and Lutz, W. (2001) An Iterative EPID Calibration Procedure for Dosimetric Verification That Considers the EPID Scattering Factor. Medical Physics, 28, 2247-2257.
http://dx.doi.org/10.1118/1.1410122

[9]   McDermott, L.N., Nijsten, S.M.J.J.G., Sonke, J.J., Partridge, M., Van Herk, M. and Mijnheer, B.J. (2006) Comparison of Ghosting Effects for Three Commercial a-Si EPIDs. Medical Physics, 33, 2448-2451.
http://dx.doi.org/10.1118/1.2207318

[10]   Fidanzio, A., Cilla, S., Greco, F., Gargiulo, L., Azario, L., Sabatino, D., et al. (2011) Generalized EPID Calibration for in Vivo Transit Dosimetry. Physica Medica, 27, 30-38.
http://dx.doi.org/10.1016/j.ejmp.2010.02.002

[11]   Van Esch, A., Depuydt, T. and Huyskens, D.P. (2004) The Use of an aSi-Based EPID for Routine Absolute Dosimetric Pre-Treatment Verification of Dynamic IMRT Fields. Radiotherapy and Oncology, 71, 223-234.
http://dx.doi.org/10.1016/j.radonc.2004.02.018

[12]   El-Mohri, Y., Antonuk, L.E., Yorkston, J., Jee, K.W., Maolinbay, M., Lam, K.L., et al. (1999) Relative Dosimetry Using Active Matrix Flat-Panel Imager (AMFPI) Technology. Medical Physics, 26, 1530-1540.
http://dx.doi.org/10.1118/1.598649

[13]   Zhu, Y., Jiang, X.Q. and Van Dyk, J. (1995) Portal Dosimetry Using a Liquid Ion Chamber Matrix: Dose Response Studies. Medical Physics, 22, 1101-1106.
http://dx.doi.org/10.1118/1.597502

[14]   Mohammadi, M. and Bezak, E. (2006) Two-Dimensional Transmitted Dose Measurements Using a Scanning Liquid Ionization Chamber EPID. Physics in Medicine and Biology, 51, 2971-2985.
http://dx.doi.org/10.1088/0031-9155/51/11/019

[15]   Alva, H., Mercado-Uribe, H., Rodriguez-Villafuerte, M. and Brandan, M.E. (2002) The Use of a Reflective Scanner to Study Radiochromic Film Response. Physics in Medicine and Biology, 47, 2925-2933.
http://dx.doi.org/10.1088/0031-9155/47/16/308

[16]   Hartmann, B., Martisikova, M. and Jakel, O. (2010) Homogeneity of Gafchromic EBT2 Film. Medical Physics, 37, 1753-1756.
http://dx.doi.org/10.1118/1.3368601

[17]   Alves, V., Cardoso, S. and da Silva, A. (2013) Gafchromic EBT2 Dosimetry via Robust Optimization. Computer Physics Communications, 184, 1708-1716.
http://dx.doi.org/10.1016/j.cpc.2013.02.023

[18]   Chytyk-Praznik, K., VanUytven, E., vanBeek, T.A., Greer, P.B. and McCurdy, B.M. (2013) Model-Based Prediction of Portal Dose Images during Patient Treatment. Medical Physics, 40, Article ID: 031713.
http://dx.doi.org/10.1118/1.4792203

[19]   Royer, P., Marchesi, V., Rousseau, V., Buchheit, I., Wolf, D., Peiffert, D. and Noel, A. (2014) Evaluation of Transit in Vivo Dosimetry Using Portal Imaging and Comparison with Measurements Using Diodes. Cancer/Radiotherapie, 18, 183-190.
http://dx.doi.org/10.1016/j.canrad.2014.03.006

[20]   Gimeno, J., Pujades, M.C., Garcia, T., Carmona, V., Lliso, F., Palomo, R., et al. (2014) Commissioning and Initial Experience with a Commercial Software for in Vivo Volumetric Dosimetry. Physica Medica, 30, 954-959.
http://dx.doi.org/10.1016/j.ejmp.2014.06.004

[21]   Piermattei, A., Fidanzio, A., Azario, L., Grimaldi, L., D’Onofrio, G., Cilla, S., et al. (2007) Application of a Practical Method for the Isocenter Point in Vivo Dosimetry by a Transit Signal. Physics in Medicine and Biology, 52, 5101-5117.
http://dx.doi.org/10.1088/0031-9155/52/16/026

[22]   Francois, P., Boissard, P., Berger, L. and Mazal, A. (2011) In Vivo Dose Verification from Back Projection of Transit Dose Measurement on the Central Axis of Photon Beams. Physica Medica, 27, 1-10.
http://dx.doi.org/10.1016/j.ejmp.2010.06.002

 
 
Top