JST  Vol.4 No.4 , December 2014
Iron-Selective Electrode Based on Phosphorylated Calix-6-Arene Derivative
ABSTRACT
The construction and performance characteristics of a novel Iron(III) membrane sensor based on (37-bis-[(diethoxy-thiophosphoryl)-oxy]-5,11,17,23,29,35-hexa-kis-(1,1-dimethylethyl)-calix[6]-arene-38,39,40,41,42-pentol) (A) were reported in this paper. The sensor was prepared by incorporating compound (A) into a plasticized polyvinyl chloride PVC membrane. The electrode reveals a Nernstian behavior over a wide iron ion concentration range (1.0 × 10-2 - 1.0 × 10-4 mol·L-1), and relatively low detection limit (1.0 × 10-5 mol·L-1). The potentiometric response was independent on the pH of the solution in the range of 4 - 7. The electrode showed a very short response time (<5 s). It exhibited very good selectivity relative to a wide variety of metal cations. Correlation of selectivity coefficient values with ionic radii of interferent cations was discussed. The proposed iron- ISE was applied for analysis of iron containing samples with comparison to a recommended spectrophotometric method.

Cite this paper
Zareh, M. , Zordek, W. and Abd-Alhady, A. (2014) Iron-Selective Electrode Based on Phosphorylated Calix-6-Arene Derivative. Journal of Sensor Technology, 4, 186-194. doi: 10.4236/jst.2014.44018.
References
[1]   Zamani, H.A., Anjali, G.M.R., Behmadi, H. and Behnajady, M.A. (2009) Fabrication of an Iron(III) PVC-Membrane Sensor Based on Bis-Benzilthiocarbohydrazide as a Selective Sensing Material. Materials Science and Engineering: C, 29, 1535.
http://dx.doi.org/10.1016/j.msec.2008.12.004

[2]   Zamani, H.A., Hamed-Mosavian, M.T., Hamidfar, E., Ganjali M.R. and Norouzi, P. (2008) A Novel Iron(III)-PVC Membrane Potentiomeric Sensor Based on N-(2-Hydroxyethyl)Ethylenediamine-N,N’,N”-Triacetic Acid. Materials Science and Engineering: C, 28, 1551.
http://dx.doi.org/10.1016/j.msec.2008.04.013

[3]   Zareh, M.M., Ismail, F.A.I. and Abd Alaziz, M.H. (2010) 1-Pyrrolidine Dicarbodithioate Based Electrodes for Determination of Iron(III) in Lubrication Oil. Electroanalysis, 22, 1369.
http://dx.doi.org/10.1002/elan.200900550

[4]   Ekmekci, G., Uzun, D., Somer, G. and Kalayci, S. (2007) A Novel Iron(III) Selective Membrane Electrode Based on Benzo-18-Crown-6 Crown Ether and Its Applications. Journal of Membrane Science, 288, 36.
http://dx.doi.org/10.1016/j.memsci.2006.10.044

[5]   Gupta, V.K., Jain, A.K., Agarwal, S. and Maheshwari, G. (2007) An Iron(III) Ion-Selective Sensor Based on a μ-bis(Tridentate) Ligand. Talanta, 71, 1964.
http://dx.doi.org/10.1016/j.talanta.2006.08.038

[6]   Sil, A., Ijeri, V.S. and Srivastava, A.K. (2005) Coated-Wire Iron(III) Ion-Selective Electrode Based on Iron Complex of 1,4,8,11-Tetraazacyclotetradecane. Sensors and Actuators B: Chemical, 106, 648.
http://dx.doi.org/10.1016/j.snb.2004.09.013

[7]   Mashhadizadeh, M.H., Shoaei, I.S. and Monadi, N. (2004) A Novel Ion Selective Membrane Potentiometric Sensor for Direct Determination of Fe(III) in the Presence of Fe(II). Talanta, 64, 1048.
http://dx.doi.org/10.1016/j.talanta.2004.05.005

[8]   Mahmoud, W.H. (2001) Iron Ion-Selective Electrodes for Direct Potentiometry and Potentiotitrimetry in Pharmaceuticals. Analytica Chimica Acta, 436, 199.
http://dx.doi.org/10.1016/S0003-2670(01)00892-3

[9]   Hassan, S.S., Mahmoud, W.H. and Othman, A.H. (1998) Ribonucleic Acid as a Novel Ionophore for Potentiometric Membrane Sensors of Some Transition Metal Ions. Talanta, 47, 377.
http://dx.doi.org/10.1016/S0039-9140(98)00142-8

[10]   Hassan, S.S.M. and Marzouk, S.A.M. (1994) A Novel Ferroin Membrane Sensor for Potentiometric Determination of Iron. Talanta, 41, 891.
http://dx.doi.org/10.1016/0039-9140(94)E0042-P

[11]   Wroblewski, W. and Brzozka, Z. (1996) Lead versus Cadmium Selectivity of Ion Selective Electrodes Based on Thiophosphorylated Calix[6]Areneionophores. New Journal of Chemistry, 20, 419.

[12]   Buck, R.P. and Lindner, E. (1994) Recommendations for Nomenclature of Ionselective Electrodes (IUPAC Recommendations 1994). Pure and Applied Chemistry, 66, 2527.
http://dx.doi.org/10.1351/pac199466122527

[13]   Eaton, A.D., Clescri, L.S. and Greenberg, A.E. (1995) Standard Methods for Examination of Water and Waste Water. American Public Health Association, Washington DC.

[14]   Zareh, M.M. (2009) Blank Membranes versus Ionophore-Based Membranes for the Selective Determination of H+. Analytical Sciences, 25, 1131.
http://dx.doi.org/10.2116/analsci.25.1131

[15]   Shannon, R.D. (1976) Revised Effective Ionic Radii and Systematic Studies of Interatomic Distances in Halides and Chalcogenides. Acta Crystallographica, A32, 751.
http://dx.doi.org/10.1107/S0567739476001551

 
 