Detection of Stiff Nodules Embedded in Soft Tissue Phantoms, Mimicking Cancer Tumours, Using a Tactile Resonance Sensor
ABSTRACT
Background: Prostate cancer (PCa) is the most common form of cancer among males in Europe and in the USA and the most common curative treatment is removal of the prostate, i.e. prostatectomy. After the removal, the prostate is histopathologically analysed. One area of interest is to examine the perifery of the prostate, as tumours on and near the surface can indicate that the PCa has spread to other parts of the body. There are no current methods to examine the surface of the prostate at the time of surgery. Tactile resonance sensors can be used for detecting areas of different stiffness in soft tissue. Human prostate tissue affected by cancer is usually stiffer than healthy tissue, and for this purpose, a tactile resonance sensor was developed. The aim of this study was to investigate the depth at which embedded stiffer volumes could be detected, using soft tissue phantoms. Methods: With the tactile resonance sensor used in this study, the shift of the resonance frequency and the force at contact with tissue can be measured, and combined into a tissue stiffness parameter. The detection sensitivity of the sensor at impression depths, 0.4 and 0.8 mm, was measured for detection of inserted nodules of stiff silicone in softer silicone and in chicken muscle tissue, mimicking prostate tissue with cancer tumours. Results: Measurements on the silicone samples detected the hidden stiffer object at a depth of 1 - 4 mm with a difference in the stiffness parameter of 80 - 900 mN/kHz (p < 0.028, n = 48). At the depth 5 - 6 mm the difference was smaller but still significant < 30 mN/kHz (p < 0.05, n = 24). For the measurements on chicken muscle, the detectable depth was 4 mm (p < 0.05, n = 24). Conclusion: This model study suggests that, with only a small impression depth of ≤1 mm, the resonance sensor system described here can detect stiffness variations located at least 4 mm in silicone and chicken muscle, mimicking tumours in prostate tissue.

Cite this paper
Åstrand, A. , Jalkanen, V. , Andersson, B. and Lindahl, O. (2014) Detection of Stiff Nodules Embedded in Soft Tissue Phantoms, Mimicking Cancer Tumours, Using a Tactile Resonance Sensor. Journal of Biomedical Science and Engineering, 7, 181-193. doi: 10.4236/jbise.2014.74022.
References
[1]   American Cancer Society (2013) http://www.cancer.org

[2]   The National Board of Health and Welfare (Socialstyrelsen) (2013) http://www.socialstyrelsen.se

[3]   Schröder, F.H., Hugosson, J., Roobol, M.J., Tammela, T.L.J., Ciatto, S., Nelen, V., Kviatkowski, M., Lujan, M., Lilja, H., Zappa, M., Denis, L.J., Recker, F., Berenguer, A., Määttänen, L., Bangma, C.H., Aus, G., Villers, A., Rebillard, X., van der Kwast, T., Blijenberg, B.G., Moss, S.M., de Koning, H.J., Auvinen, A. and E.R.S.P.C Investigators (2009) Screening and Prostate-Cancer Mortality in Randomized European Study. New England Journal of Medicine, 360, 1320-1328.
http://dx.doi.org/10.1056/NEJMoa0810084

[4]   Candefjord, S., Ramser, K. and Lindahl, O.A. (2009) Technologies for Localization and Diagnosis of Prostate Cancer. Journal of Medical Engineering & Technology, 33, 585-683.

[5]   Wang, J., Wang, J., Sun, X., Song, W., Nor, J.E., Wang, C.Y. and Taichman, R. (2005) Diverse Signalling Pathways through the SDF-1/CXCR4 Chemokine Axix in Prostate Cancer Cell Lines Leads to Altered Patterns of Cytokine Secretion and Angiogenesis. Cellular Signalling, 17, 1578-1592.
http://dx.doi.org/10.1016/j.cellsig.2005.03.022

[6]   Lindahl, O.A., Constantinou, C.E., Eklund, A., Murayama, Y., Hallberg, P. and Omata, S. (2009) Tactile Resonance Sensors in Medicine. Journal of Medical Engineering & Technology, 33, 263-273.
http://dx.doi.org/10.1080/03091900802491188

[7]   Gwilliam, J.C., Pezzementi, Z., Jantho, E., Okamura, A.M. and Hsiao, S. (2010) Human vs. Robotic Tactile Sensing: Detecting Lumps in Soft Tissue. IEEE Haptics Symposium 2010, Waltham, 21-28.

[8]   Tiwana, M.I., Redmond, S.J. and Lovell, N.H. (2012) A Review of Tactile Sensing Technologies with Applications in Biomedical Engineering. Sensors and Actuators, A179, 17-31.
http://dx.doi.org/10.1016/j.sna.2012.02.051

[9]   Jalkanen, V., Andersson, B.M., Bergh, A., Ljungberg, B. and Lindahl, O.A. (2006) Resonance Sensor Measurements of Stiffness Variations in Prostate Tissue in Vitro—A Weighted Tissue Proportion Model. Physiological Measurement, 27, 1373-1386.
http://dx.doi.org/10.1088/0967-3334/27/12/009

[10]   Krouskop, T.A., Wheeler, T.M., Kallel, F., Garra, B.S. and Hall, T. (1998) Elastic Moduli of Breast and Prostate Tissues under Compression. Ultrason Imaging, 20, 260-274.
http://dx.doi.org/10.1177/016173469802000403

[11]   Sakai, N., Tatsuta, M., Yano, H., Iishi, H. and Ishiguro, S. (2000) Diagnosis of the Extent of Gastric Cancers by a New Endoscopic Ultrasonic Tactile Sensor. Gastrointest Endosc, 51, 69-73.
http://dx.doi.org/10.1016/S0016-5107(00)70390-7

[12]   Lindahl, O.A., Ängquist, K.-A. and Ödman, S. (1991) Impression Technique for the Assessment of Oedema—Technical Improvement and Methodological Evaluation of a New Technique. Medical & Biological Engineering & Computing, 29, 591-597. http://dx.doi.org/10.1007/BF02446090

[13]   Sasai, S., Zhen, Y.-X., Suetake, T., Tanita, Y., Omata, S. and Tagami, H. (1999) Palpation of the Skin with a Robot Finger: An Attempt to Measure Skin Stiffness with a Probe Loaded with a Newly Developed Tactile Vibration Sensor and Displacement Sensor. Skin Research Technology, 5, 237-246.
http://dx.doi.org/10.1111/j.1600-0846.1999.tb00136.x

[14]   Miyaji, K., Furuse, A., Nakajima, J., Kohno, T., Ohtsuka, T., Yagyu, K., Oka, T. and Omata, S. (1997) The Stiffness of Lymph Nodes Containing Lung Carcinoma Metastases—A New Diagnostic Parameter Measured by a Tactile Sensor. Cancer, 80, 1920-1925.
http://dx.doi.org/10.1002/(SICI)1097-0142(19971115)80:10<1920::AID-CNCR8>3.0.CO;2-R

[15]   Kusaka, K., Harihara, Y., Torzilli, G., Kubota, K., Takayama, T., Makuuchi, M., Mori, M. and Omata, S. (2000) Objective Evaluation of Liver Consistency to Estimate Hepatic Fibrosis and Functional Reserve for Hepatectomy. Journal of the American College of Surgeons, 191, 47-53.
http://dx.doi.org/10.1016/S1072-7515(00)00309-4

[16]   Jalkanen, V., Andersson, B.M., Bergh, A., Ljungberg, B. and Lindahl, O.A. (2006) Prostate Tissue Stiffness as Measured with a Resonance Sensor System: A Study on Silicone and Human Prostate Tissue in Vitro. Medical & Biological Engineering & Computing, 44, 593-603.
http://dx.doi.org/10.1007/s11517-006-0069-6

[17]   Åstrand, A.P., Jalkanen, V., Andersson, B.M. and Lindahl, O.A. (2013) Contact Angle and Indentation Velocity Dependency for a Resonance Sensor—Evaluation on Soft Tissue Silicone Models. Journal of Medical Engineering & Technology, 37, 185-196.
http://dx.doi.org/10.3109/03091902.2013.773097

[18]   Åstrand, A.P. (2012) A Flexible Resonance Sensor System for Detection of Cancer Tissue—Evaluation on Silicone. Lic. Thesis, Chapter 4, 5-8. Umeå University, Umeå.

[19]   Jalkanen, V., Andersson, B.M., Bergh, A., Ljungberg, B. and Lindahl, O.A. (2008) Explanatory Models for a Tactile Resonance Sensor System—Elastic and Density-Related Variations of Prostate Tissue in Vitro. Physiological Measurement, 29, 729-745.
http://dx.doi.org/10.1088/0967-3334/29/7/003

[20]   Jalkanen, V., Andersson, B.M. and Lindahl, O.A. (2010) Stiffness of a Small Tissue Phantom Measured by a Tactile Resonance Sensor. Proceedings of XII Mediterranean Conference on Medical and Biological Engineering and Computing (Chalkidiki) IFMBE Proceedings, 29, 395-398.

[21]   Omata, S. and Terunuma, Y. (1992) New Tactile Sensor like the Human Hand and Its Applications. Sensors and Actuators, 35, 9-15.
http://dx.doi.org/10.1016/0924-4247(92)87002-X

[22]   Wacker-Chemie GmbH (2011) http://www.wacker.com

[23]   Eklund, A., Bergh, A. and Lindahl, O.A. (1999) A Catheter Tactile Sensor for Measuring Hardness of Soft Tissue: Measurement in a Silicone Model and in an in Vitro Human Prostate Model. Medical & Biological Engineering & Computing, 37, 618-624.
http://dx.doi.org/10.1007/BF02513357

[24]   Lindahl, O.A. and Omata, S. (1995) Impression Technique for the Assessment of Oedema: Comparison with a New Tactile Sensor That Measures Physical Properties of Tissue. Medical & Biological Engineering & Computing, 33, 27-32.
http://dx.doi.org/10.1007/BF02522941

[25]   Beuth Verlag GmbH (2013) http://www.beuth.de

[26]   American Society for Testing and Materials, ASTM (2013)
http://www.astm.org/Standards/D2240.htm

[27]   Murayama, Y. (2005) Development of a Micro Tactile Sensor and Its Biomedical Applications. PhD Thesis, Chapter 5. Nihon University, Nihon.

[28]   Kim, Y., Ahn, B., Lee, J.W., Rha, K.H. and Kim, J. (2013) Local Property Characterization of Prostate Glands Using Inhomogeneous Modelling Based on Tumor Volume and Location Analysis. Medical & Biological Engineering & Computing, 51, 197-205.
http://dx.doi.org/10.1007/s11517-012-0984-7

[29]   Sangpradit, K., Liu, H., Dasgupta, P., Althoefer, K. and Seneviratne, L.D. (2011) Finite-Element Modelling of Soft Tissue Rolling Indentation. IEEE Transactions on Biomedical Engineering, 58, 3319-3327.
http://dx.doi.org/10.1109/TBME.2011.2106783

[30]   Liu, H., Sangpradit, K., Li, M., Dasgupta, P., Althoefer, K. and Seneviratne, L.D. (2014) Inverse Finite-Element Modeling for Tissue Parameter Identification Using a Rolling Indentation Probe. Medical & Biological Engineering & Computing, 52, 17-28.
http://dx.doi.org/10.1007/s11517-013-1118-6

[31]   Bergström, J.S. and Boyce, M.C. (1998) Constitutive Modelling of the Large Strain Time-Dependent Behaviour of Elastomers. Journal of the Mechanics and Physics of Solids, 46, 931-954.
http://dx.doi.org/10.1016/S0022-5096(97)00075-6

[32]   Martins, P.A.L.S., Natal, J.R.M. and Ferreira, A.J.M. (2006) A Comparative Study of Several Material Models for Prediction of Hyperelastic Properties: Application to Silicone-Rubber and Soft Tissue. Strain, 42, 135-147.
http://dx.doi.org/10.1111/j.1475-1305.2006.00257.x

[33]   Fung, Y.C. (1993) Bioviscoelastic Solids. In Biomechanics-Mechanical Properties of Living Tissues. 2nd Edition, Springer-Verlag Inc., New York, 242-311.

[34]   Jalkanen, V., Andersson, B.M., Bergh, A., Ljungberg, B. and Lindahl, O.A. (2007) Spatial Variations in Prostate Tissue Histology Measured by a Tactile Resonance Sensor. Physiological Measurement, 228, 1267-1281.
http://dx.doi.org/10.1088/0967-3334/28/10/011

 
 
Top