Is It Possible to Publish a Calibration Function for Radiochromic Film?
ABSTRACT

Purpose: To assess the possibility of using a public calibration function for radiochromic film dosimetry in dose QA of highly conformal treatment plans. Methods: EBT3 film calibration strips (3.5 × 20 cm2 from lots A101212 and A011713) were exposed on a Varian Trilogy at a facility to a 10 × 10 cm2 open field at doses of 80, 160, 320 cGy using 6MV photons. Together with a strip of unexposed film from the same lot, the exposed films were digitized in a single scan using different Epson 10,000 XL scanners at two different facilities. The dose-response data for each color-channel from each facility were generated using the same calibration function X(D) = a + b/(D - c), where X(D) is the response at dose D and a, b and c are the coefficients. Different batches of EBT3 film were exposed to a VMAT beam. These films, plus two reference strips exposed to doses of zero and 160 cGy, were digitized on the scanners at the two facilities. Using the multi-channel dosimetry method and One-scan protocol (Med Phys, 39: 6339-6349, 2012) the recorded doses on the VMAT films were calculated and the results were compared with the VMAT plan using a Gamma index of 3%/3 mm. Results: The passing rates obtained for dose maps calculated for all combinations of VMAT images and calibration functions were nearly unchanged, using the One-scan protocol. Also, in all cases a passing rate of >99% was obtained for Gamma index of 3%/3 mm. On the other hand, if the One-scan protocol was not employed, the dose maps for VMAT images and calibration functions from different scanners showed poor correlation with the treatment plan. This is probably due to the scan-to-scan variability. Conclusions: The authors have found that it is feasible to use a public calibration function for a given radiochromic film lot using the same methodology, One-scan protocol, for patient-specific QA.


Cite this paper
M. Chan, D. Lewis and X. Yu, "Is It Possible to Publish a Calibration Function for Radiochromic Film?," International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol. 3 No. 1, 2014, pp. 25-30. doi: 10.4236/ijmpcero.2014.31005.
References
[1]   M. F. Chan, Q. Zhang, J. Li, P. Parhar, K. Schupak and C. Burman, “The Verification of iPlan Commissioning by Radiochromic EBT2 Films,” International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol. 1, No. 1, 2012, pp. 1-7.
http://dx.doi.org/10.4236/ijmpcero.2012.11001

[2]   C. G. Soares, S. Trichter and S. Devic, “Radiochromic Film Dosimetry,” In: D. W. O. Rogers and J. Cygler, Eds., Clinical Dosimetry for Radiotherapy, Medical Physics Publishing, Madison, 2009, pp. 759-813.

[3]   S. Chiu-Tsao, D. Medich and J. Munro, “The Use of New GAFCHROMIC® EBT Film for 125I Seed Dosimetry in Solid Water® Phantom,” Medical Physics, Vol. 35, No. 8, 2008, pp. 3787-3799.
http://dx.doi.org/10.1118/1.2955746

[4]   M. Todorovic, M. Fischer, F. Cremers, E. Thom and R. Schmidt, “Evaluation of GafChromic EBT Prototype B for External Beam Dose Verification,” Medical Physics, Vol. 33, No. 5, 2006, pp. 1321-1328.
http://dx.doi.org/10.1118/1.2188077

[5]   F. C. Su, Y. Liu, S. Stathakis, C. Shi, C. Esquivel and N. Papanikolaou, “Dosimetry Characteristics of GAFCHROMIC® EBT Film Responding to Therapeutic Electron Beams,” Applied Radiation and Isotopes, Vol. 65, No. 10, 2007, pp. 1187-1192.
http://dx.doi.org/10.1016/j.apradiso.2007.05.005

[6]   S. Chiu-Tsao and M. F. Chan, “Photon Beam Dosimetry in the Superficial Buildup Region Using Radiochromic EBT Film Stack,” Medical Physics, Vol. 36, No. 6, 2009, pp. 2074-2083. http://dx.doi.org/10.1118/1.3125134

[7]   A. Niroomand-Rad, C. R. Blackwell, B. M. Coursey, K. P. Gall, J. M. Galvin, W. L. McLaughlin, A. S Meigooni, R. Nath, J. E. Rodger and C. G. Soares, “Radiochromic Film Dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55,” Medical Physics, Vol. 25, No. 11, 1998, pp. 2093-2115.
http://dx.doi.org/10.1118/1.598407

[8]   S. Devic, J. Seuntjens, W. Abdel-Rahman, M. Evans, M. Olivares, E. B. Podgorsak, T. Vuong and C. G. Soares, “Skin Dose Measurement Using Radiochromic Film in Clinical Applications,” Medical Physics, Vol. 33, No. 4, 2006, pp. 1116-1124.
http://dx.doi.org/10.1118/1.2179169

[9]   S. Price, M. Williams, M. Butson and P. Metcalfe, “Comparison of Skin Dose between Conventional Radiotherapy and IMRT,” Australasian Physical and Engineering Sciences in Medicine, Vol. 29, No. 3, 2006, pp. 272-277. http://dx.doi.org/10.1007/BF03178577

[10]   H. Chung, H. Jin, J. F. Dempsey, C. Liu, J. Palta, T.-S. Suh and S. Kim, “Evaluation of Surface and Build-Up Region Dose for Intensity-Modulated Radiation Therapy in Head and Neck Cancer,” Medical Physics, Vol. 32, No. 8, 2005, pp. 2682-2689.
http://dx.doi.org/10.1118/1.1992067

[11]   S. Chiu-Tsao and M. F. Chan, “Evaluation of Two-Dimensional Bolus Effect of Immobilization/Support Devices on Skin Doses: A Radiochromic EBT Film Dosimetry Study in Phantom,” Vol. 37, No. 7, 2010, pp. 3611-3626. http://dx.doi.org/10.1118/1.3439586

[12]   L. J. van Battum, D. Hoffmans, H. Piersma and S. Heukelom, “Accurate Dosimetry with Gafchromic EBT Film of a 6 MV Photon Beam in Water: What Level Is Achievable?” Medical Physics, Vol. 35, No. 2, 2008, pp. 704-716. http://dx.doi.org/10.1118/1.2828196

[13]   C. Soares, “New Developments in Radiochromic Film Dosimetry,” Radiation Protection Dosimetry, Vol. 20, No. 1-4, 2006, pp. 100-106.
http://dx.doi.org/10.1093/rpd/nci698

[14]   S. Devic, N. Tomic, C. G. Soares and E. B. Podgorsak, “Optimizing the Dynamic Range Extension of a Radiochromic Film Dosimetry System,” Medical Physics, Vol. 36, No. 2, 2009, pp. 429-437.
http://dx.doi.org/10.1118/1.3049597

[15]   J. E. Lye, D. J. Butler and D. V. Webb, “Enhanced Epidermal Dose Caused by Localized Electron Contamination from Lead Cutouts Used in Kilovoltage Radiotherapy,” Medical Physics, Vol. 37, No. 8, 2010, pp. 3935-3939. http://dx.doi.org/10.1118/1.3458722

[16]   M. Oldham, H. Sakhalkar, P. Guo and J. Adamovics, “An Investigation of the Accuracy of an IMRT Dose Distribution Using Two- and Three-Dimensional Dosimetry Techniques,” Medical Physics, Vol. 35, No. 5, 2008, pp. 2072-2080. http://dx.doi.org/10.1118/1.2899995

[17]   D. Fontanarosa, L. C. Orlandini, I. Andriani and L. Bernardi, “Commissioning Varian Enhanced Dynamic Wedge in the Pinnacle Treatment Planning System Using GafchromicTM EBT Film,” Medical Physics, Vol. 36, No. 10, 2009, pp. 4504-4510.
http://dx.doi.org/10.1118/1.3223621

[18]   J. Novotny Jr., J. P. Bhatnagar, M. A. Quader, G. Bednarz, L. D. Lunsford and M. S. Huq, “Measurement of Relative Output Factors for the 8 and 4 mm Collimators of Leksell Gamma Knife Perfexion by Film Dosimetry,” Medical Physics, Vol. 36, No. 5, 2009, pp. 1768-1774.
http://dx.doi.org/10.1118/1.3113904

[19]   M. F. Chan, S. Chiu-Tsao, J. Li, K. Schupak, P. Parhar and C. Burman, “Confirmation of Skin Doses Resulting from Bolus Effect of Intervening Alpha-Cradle and Carbon Fiber Couch in Radiotherapy,” Technology in Cancer Research and Treatment, Vol. 11, No. 6, 2012, pp. 571-581.

[20]   S. Chiu-Tsao and M. F. Chan, “Dose Response Characteristics of the New EBT-2 Film for Different Megavoltage Beams,” Medical Physics, Vol. 36, No. 6, 2009, p. 2604.
http://dx.doi.org/10.1118/1.3181842

[21]   M. Fragoso, N. Wen, S. Kumar, D. Liu, S. Ryu, B. Movsas, M. Ajlouni and I. J. Chetty, “Dosimetric Verification and Clinical Evaluation of a New Commercially Available Monte Carlo-Based Dose Algorithm for Application in Stereotactic Body Radiation Therapy (SBRT) Treatment Planning,” Physics in Medicine and Biology, Vol. 55, No. 16, 2010, pp. 4445-4464.
http://dx.doi.org/10.1088/0031-9155/55/16/S02

[22]   O. A. Zeidan, S. A. L. Stephenson, S. L. Meeks, T. H. Wagner, T. R. Willoughby, P. A. Kupelian and K. M. Langen, “Characterization and Use of EBT Radiochromic Film for IMRT Dose Verification,” Medical Physics, Vol. 33, No. 11, 2006, pp. 4064-4072.
http://dx.doi.org/10.1118/1.2360012

[23]   S. Devic, J. Seuntjens, G. Hegyi, E. Podgorsak, C. Soares, A. Kirov, I. Ali, J. Williamson and A. Elizondo, “Dosimetric Properties of Improved Gafchromic Films for Seven Different Digitizers,” Medical Physics, Vol. 31, No. 9, 2004, pp. 2391-2398.
http://dx.doi.org/10.1118/1.1776691

[24]   “Gafchromic® EBT2: Self-Developing Film for Radiotherapy Dosimetry,” ISP White Paper, 2010.
http://www.filmqapro.com/Documents/GafChromic_EBT-2_20101007.pdf

[25]   E. Willcox, G. Daskalov and L. Nedialkova, “Comparison of the Epson Expression 1680 Flatbed and the VXR-16 Dosimetry PROTM Film Scanners for use in IMRT Dosimetry Using Gafchromic and Radiographic Film,” Medical Physics, Vol. 34, No. 1, 2007, pp. 41-48.
http://dx.doi.org/10.1118/1.2402584

[26]   A. Micke, D. F. Lewis and X. Yu, “Multichannel Film Dosimetry with Nonuniformity Correction,” Medical Physics, Vol. 38, No. 5, 2011, pp. 2523-2534.
http://dx.doi.org/10.1118/1.3576105

[27]   D. Lewis, A. Micke, X. Yu and M. F. Chan, “An Efficient Protocol for Radiochromic Film Dosimetry Combining Calibration and Measurement in a Single Scan,” Medical Physics, Vol. 39, No. 10, 2012, pp. 6339-6350.
http://dx.doi.org/10.1118/1.4754797

[28]   S. Devic, N. Tomic, S. Aldelaijan, F. DeBlois, J. Seuntjens, M. F. Chan and D. Lewis, “Linearization of Dose Response Curve of the Radiochromic Film Dosimetry System,” Medical Physics, Vol. 39, No. 8, 2012, pp. 4850-4857. http://dx.doi.org/10.1118/1.4736800

[29]   M. J. Butson, T. Cheung and P. K. Yu, “Weak Energy Dependence of EBT Gafchromic Film Dose Response in the 50 kVp-10 MVp X-Ray Range,” Applied Radiation and Isotopes, Vol. 64, No. 1, 2006, pp. 60-62.
http://dx.doi.org/10.1016/j.apradiso.2005.07.002

[30]   S. Chiu-Tsao, Y. Ho, R. Schankar, L. Wang and L. B. Harrison, “Energy Dependence of Response of New High Sensitivity Radiochromic Films for Megavoltage and Kilovoltage Radiation Energies,” Medical Physics, Vol. 32, No. 11, 2005, pp. 3350-3354.
http://dx.doi.org/10.1118/1.2065467

[31]   M. J. Butson, P. K. N. Yu, T. Cheung and H. Alnawaf, “Energy Response of the New EBT2 Radiochromic Film to X-Ray Radiation,” Radiation Measurements, Vol. 45, No. 7, 2010, pp. 836-839.
http://dx.doi.org/10.1016/j.radmeas.2010.02.016

[32]   B. Arjomandy, R. Tailor, A. Anand, N. Sahoo, M. Gillin, K. Prado and M. Vicic, “Energy Dependence and Dose Response of Gafchromic EBT2 Film over a Wide Range of Photon, Electron, and Proton Beam Energies,” Medical Physics, Vol. 37, No. 5, 2010, pp. 1942-1947.
http://dx.doi.org/10.1118/1.3373523

[33]   P. Lindsay, A. Rink, M. Ruschin and D. Jaffray, “Investigation of Energy Dependence of EBT and EBT-2 Gafchromic Film,” Medical Physics, Vol. 37, No. 2, 2010, pp. 571-576. http://dx.doi.org/10.1118/1.3291622

[34]   J. G. H. Sutherland and D. W. O. Rogers, “Monte Carlo Calculated Absorbed-Dose Energy Dependence of EBT and EBT2 Film,” Medical Physics, Vol. 37, No. 3, 2010, pp. 1110-1116. http://dx.doi.org/10.1118/1.3301574

[35]   G. Massillon-JL, S. Chiu-Tsao, I. Domingo-Munoz and M. F. Chan, “Energy Dependence of the New Gafchromic EBT3 Film: Dose Response Curves for 50 kV, 6 and 15 MV X-Ray Beams,” International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol. 1, No. 3, 2012, pp. 60-65.
http://dx.doi.org/10.4236/ijmpcero.2012.12008

 
 
Top